12.7
Process connection label (pressure rated versions)
For pressure-rated versions only, the process connection label lists the following information:
SITRANS LR560 (mA/HART)
Operating Instructions, 06/2017, A5E34647946-AD
Item Sample Text
CONNECTION SERIES ASME B16.5 / EN 1092-1/ JIS B
NOM. PIPE SIZE (DN) 4 INCH / 100mm
MAWP (PS) 3 BAR
DESIGN TEMP. (TS) 100 ºC
MIN. PROCESS 3 BAR AT -40 ºC
0F13589.5
TEST PRESSURE (PT) 5.2 BAR
TEST DATE 10/01/04
PROCESS SERIES 25785
WETTED NON-METALLIC PEI
WETTED METALLICS 304L
WETTED SEALS FKM / VQM
12.7 Process connection label (pressure rated versions)
2220
Dimension drawings
Comments/Explanation
Flange Series: dimensional
pattern based on ASME
B16.5/EN 1092-1/JIS B 2 220
flange standards
Nominal Pipe Size: based on
150#/PN16/10K flange pressure
classes
Maximum Allowable Working
Pressure at Design Tempera-
ture
Maximum Allowable Working
Temperature
Minimum Wetted Process Con-
ditions
Canadian Registration Number
(CRN)
Production Test Pressure
Date of Pressure Test
(Year/Month/Day)
Pressure Tag Family Series
Sensor Lens Material
Process Connection Material(s)
Seal Material(s)
177