6.2. Local Access
6.2.1. Write Protection
Prevents any changes to parameters via remote or local access.
Options
•
To turn Lock On, key in any value other than the Unlock Value.
•
To turn Lock Off, key in the Unlock Value.
To access this parameter via AMS Device Manager see Write Protection on
page 67 under
6.2.2. Local Operation
Enables or disables programming via the handheld programmer.
Options
Note: Once disabled via the handheld programmer, the parameter is no
longer visible on the local display and can only be reset using AMS Device
Manager. However, if no communication activity exists for 30 seconds, the
parameter will again be visible on the local display.
To access this parameter via AMS Device Manager see Local Display on page
59 under
7. Language
Selects the language to be used on the LCD.
Options
To access this parameter via AMS Device Manager, see
Page 122
SITRANS LR560 (FF) – INSTRUCTION MANUAL
Range: 0 to 9999
*
Unlock value 2457
Any other value
Security (RESOURCE)
DISABLED
*
ENABLED
Setup (LCD)
.
* ENGLISH
DEUTSCH
FRANÇAIS
ESPAÑOL
简体中文
.
Local Display
Lock Off
Lock On
on page 59.
7ML19985LY01