Section 3.0
Interconnect Runs, Wiring and Cables
3.1
GEMS Supplied (Optional, Short Run)
LENGTH,
ACTUAL
(USABLE)
PART # DESCRIPTION
ft
m
050
28
8.5
2343529-2 HVDC, PDU to Gantry
(20)
(6.1)
051
28
8.5
2343530-2 HVAC, PDU to Gantry
(20)
(6.1)
052
28
8.5
2343528-2 LVAC, PDU to Gantry
(20)
(6.1)
053
65
19.8
2343531-2 LVAC, PDU to Console
(60)
(18.3)
054
n/a
055
28
8.5
2371450-2 Ground, PDU to Raceway
(20)
(5.97)
056
68
20.8
2371450-4 Ground, Raceway to Console 1283 VW-1
(57)
(17.4)
100 32.5
9.9
5120646-2 Signal, Gantry MSUB to PDU
(20)
(6.1)
101
71
21.7
5120645-2 Signal, Gantry MSUB to OC
(60)
(18.3)
102
71
21.7
2373436-3 Signal (LAN), Gantry to OC
(63)
(19.3)
103
68
20.7
2117848-7 Fiber Optic, Gantry to OC
(60)
(18.3)
104
n/a
Table 9-2 GEMS Supplied Cables (Optional, Short Run) - UL Information
UL CABLE INFORMATION
2587 FT4
2587 FT4
2587 FT4
2587 FT4
LVAC, Gantry to Table
1015
1284 VW-1
Signal, Gantry to Table
600
+ & - 350VDC 90
600
440Y/254
90
600
208Y/120
90
600
120VAC
90
600
120VAC
600
0
105
(FT-1)
600
0
105
(FT-1)
FT-4
300
<30VDC
80
FT-4
300
<30VDC
80
1900 <30VDC
N/A
N/A
FT-4
300
80
PULL SIZE
MM (INCHES)
19 (.751)
3
(2) 4
22 (0.87) Dia
(1) 8
15.3 (.604) 4
14
11.2 (0.44) Dia
13.8 (.542) 5
8
56.4 (2.22) Dia
12.3 (.483) 3
10
56.4 (2.22) Dia
3
14
15.5 (.608) 1
1/0
15.8 (0.62) Dia
11.9 (.467) 1
2
12.2 (0.48) Dia
11.2 (.440) 25 22
17 x 58 (0.68 x 2.30)
19 x 51 (0.75 x 2.01)
11.2 (.440) 25 22
17 x 58 (0.68 x 2.30)
19 x 51 (0.75 x 2.01)
5.9 (.234)
8
24
15 (0.59) Dia
1
N/A 10 (0.39) Dia
25 22