GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
10-7-3
Outlet Test Wiring Arrangement - USA & Canada
Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet
tester and noting the combination of lights that are illuminated. Any problems found should be reported
to the hospital immediately and the receptacle should not be used.
NOTE:
No outlet tester can detect the condition in which the Neutral (grounded supply) conductor and the
Grounding (protective earth) conductor have been reversed. If later tests indicate high current leakages,
this should be suspected as a possible cause and the outlet wiring should be visually inspected.
10-7-4
Grounding Continuity
Electric Shock Hazard!
CAUTION
The patient or operator must not come into contact with the equipment during this test.
Measure the resistance from the third pin of the attachment plug to the exposed metal parts of the case.
The Ground wire resistance should be less than 0.2 Ohms. Refer to the procedure in the IEC 601-1.1.
GROUND PIN
10-20
04
EVISION
Figure 10-1 Typical Outlet Tester
Figure 10-2 Ground Continuity Test
Section 10-7 - Electrical Safety Tests
CONSOLE
OHMMETER
V
3N P
/E
IVID
RO
XPERT
ACCESSIBLE METAL PARTS
such as:
a) Monitor Housing
b) Rear Panel Connector
c) Any Caster/Wheel support
S
M
ERVICE
ANUAL