hit counter script

Siemens SITRANS L Operating Instructions Manual page 211

Radar transmitters with profibus pa
Hide thumbs Also See for SITRANS L:
Table of Contents

Advertisement

LCD display
Meaning
S:0
The device was unable to get a meas-
urement within the Failsafe LOE Timer
period. Possible causes: faulty installa-
tion, antenna material buildup, foam-
ing/other adverse process conditions,
invalid calibration range.
S:2
Unable to collect profile because of a
power condition that is outside the oper-
ating range of the device.
S:3
Device is nearing its lifetime limit accord-
ing to the value set in Maintenance Re-
quired Limit.
S:4
Device is nearing its lifetime limit accord-
ing to the value set in Maintenance De-
manded Limit.
S:6
Sensor is nearing its lifetime limit accord-
ing to the value set in Maintenance Re-
quired Limit.
S:7
Sensor is nearing its lifetime limit accord-
ing to the value set in Maintenance De-
manded Limit.
S:8
Service interval as defined in Mainte-
nance Required Limit has expired.
S:9
Service interval as defined in
Maintenance Demanded Limit
has expired.
S:10
Input parameters High Calibration Point
and Low Calibration Point are the same.
S:11
Internal temperature sensor failure.
S:12
* Internal temperature of the device has
exceeded
specifications: it is operating outside its
temperature range.
SITRANS LR200 with PROFIBUS PA
Operating Instructions, 07/2021, A5E32337680-AF
Acyclic Extended Diagnostics /General Fault Codes
E.20 Acyclic extended diagnostics (general fault codes)
Corrective Action
Ensure installation details are correct.
Ensure no antenna material
buildup. Clean if necessary.
Adjust process conditions to minimize
foam or other adverse conditions.
Correct range calibration.
If fault persists, contact your local Sie-
mens representative.
Repair required. Contact your local Siemens
representative.
Replacement is recommended.
Replacement is recommended.
Replacement is recommended.
Replacement is recommended.
Perform service
Perform service.
Check calibration settings of device.
Ensure settings for High Calibration
Point and Low Calibration Point are dif-
ferent.
Repair required. Contact your local Siemens
representative.
Relocate device and/or lower process
temperature enough to cool device.
Inspect for heat-related damage and
contact your local Siemens representa-
tive if repair is
required.
Fault code will persist until a manual
reset is performed using PDM or the
LCD interface.
PROFIBUS communication
Byte
Bit
0
2
0
3
4
6
7
0
1
3
4
5
1
211

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sitrans lr200

Table of Contents