Page 2
Introduction Safety notes SITRANS Description Installing/mounting Sensing Heads SITRANS WFS320 Connecting Commissioning Operating Instructions Service and maintenance Diagnosing and troubleshooting Technical Data Dimension drawings Appendix A 02/2016 A5E32880290-AB...
Page 3
Note the following: WARNING Siemens products may only be used for the applications described in the catalog and in the relevant technical documentation. If products and components from other manufacturers are used, these must be recommended or approved by Siemens. Proper transport, storage, installation, assembly, commissioning, operation and maintenance are required to ensure that the products operate safely and without any problems.
This manual covers only sensing head installation, operation, and maintenance procedures. Flowmeter and integrator instruction manuals are available for download from our web site: Siemens weighing (http://www.siemens.com/weighing) Follow these operating instructions for quick, trouble-free installation, and maximum accuracy and reliability of your device.
Installation in hazardous areas WARNING Equipment used in hazardous areas must be Ex-approved and marked accordingly. It is required that the special conditions for safe use provided in the manual and in the Ex certificate are followed. SITRANS WFS320 Operating Instructions, 02/2016, A5E32880290-AB...
Page 9
IIA, IIB, and IIC, and temperature classes T1, T2, T3, T4, T5, and T6 in zone 2 hazardous locations. 4. The equipment is certified for use as Category 2D equipment per FM14ATEX0027X. SITRANS WFS320 Operating Instructions, 02/2016, A5E32880290-AB...
Page 10
● The enclosure is non-conducting and, under certain extreme conditions, may generate an ignition-capable level of electrostatic charges. The user shall ensure that the equipment is not installed in a location where it may be subjected to external conditions (such as high-pressure SITRANS WFS320 Operating Instructions, 02/2016, A5E32880290-AB...
Page 11
This shall be considered when the WFS320 is being installed in locations that specifically require Equipment Protection Level Db. SITRANS WFS320 Operating Instructions, 02/2016, A5E32880290-AB...
Description SITRANS WFS320 SITRANS WFS320 sensing head is an out-of-process sensing element used for continuous in- line weighing of powdered or granular dry bulk solid materials. It is used with the 300 tph versions of SITRANS WF330 (general purpose), WF340 (vertical material drop), and WF350 (aerated gravity conveyor) dry solids flowmeters.
Page 13
Description 3.2 Principle of operation ① ③ Impact force Moving beam travel ② ④ LVDT Sensing plate SITRANS WFS320 Operating Instructions, 02/2016, A5E32880290-AB...
Ensure that the structure used to support the base mount sensing head is capable of supporting the dynamic material impact forces as well as the static weight of the sensing head. ① ③ Spirit level Supporting/levelling bolts ② ④ Moving beam (machined surface) Rigid support SITRANS WFS320 Operating Instructions, 02/2016, A5E32880290-AB...
5. Tighten the right taper pin nut to lock the sensing plate shaft in place. 6. Tighten the left taper pin nut. ① ③ Taper pin (flat side down) Socket ② ④ Shaft Flat side up SITRANS WFS320 Operating Instructions, 02/2016, A5E32880290-AB...
The damper must be full and free of air bubbles, with the damper cover in the UP position, during flowmeter operation. ① ④ Filler bottle Damper cover ② ⑤ Filler plug Check that there are no bubbles ③ Shipping screws SITRANS WFS320 Operating Instructions, 02/2016, A5E32880290-AB...
Not applicable to Hazardous-rated units. LVDT to SF500 Integrator LVDT to LVDT conditioner card connections are made by Siemens. *Where separation between the integrator and LVDT conditioner exceeds 150 m (500 ft): 1. Remove the jumpers from SF500 terminals 11/12 and 13/14.
Page 20
5.2 Unit with sensing head mounted LVDT conditioner card LVDT to SIWAREX FTC LVDT to LVDT conditioner card connections are made by Siemens. Connect the shield to the spring clamp shielding element that is mounted below the FTC. See ♦...
– from SF500 terminal 12 to conditioner terminal block marked "Integrator EXC+" – from SF500 terminal 13 to conditioner terminal block marked "Integrator EXC-" For further connection information on specific LVDTs, consult Siemens. ** For Encapsulated LVDT (hazardous) • YEL = WHITE •...
Page 22
Shields are common, but not grounded to chassis. Run cable shields through SHLD terminals ♦ and ground at Integrator only. Note Ensure that the connection between TB-2 and TB-17 is made. Note Junction box included with hazardous-rated unit. SITRANS WFS320 Operating Instructions, 02/2016, A5E32880290-AB...
Page 23
***The LVDT can be installed in a hazardous area, but all other components must be in a non- hazardous area. Connect the shield to the spring clamp shielding element that is mounted below the FTC. See ♦ FTC manual for details. Note Junction box included with hazardous-rated unit. SITRANS WFS320 Operating Instructions, 02/2016, A5E32880290-AB...
Page 24
Connecting 5.3 Unit with remote-located LVDT conditioner card SITRANS WFS320 Operating Instructions, 02/2016, A5E32880290-AB...
Page 25
Connecting 5.3 Unit with remote-located LVDT conditioner card SITRANS WFS320 Operating Instructions, 02/2016, A5E32880290-AB...
*For WF350 flowmeters, use 60 grams (instead of 45 grams) when calculating Test Rate or Test Weight. Example If the test weight used with a WF330 flowmeter is 7000 grams: Note Use metric tons per hour (t/h) or short tons per hour (STPH) as applicable for TPH. SITRANS WFS320 Operating Instructions, 02/2016, A5E32880290-AB...
– Tighten the locknut, ensuring the measured value is maintained. ① ③ Locknut LVDT core ② ④ Moving beam LVDT Note Ensure the new position of the LVDT core allows free movement within the LVDT bore. SITRANS WFS320 Operating Instructions, 02/2016, A5E32880290-AB...
2. Remove the test weight and readjust the LVDT Output Zero, if necessary. If the level test is performed after the integrator has been calibrated, a new integrator zero and span calibration, span adjust, and factoring, should be performed. SITRANS WFS320 Operating Instructions, 02/2016, A5E32880290-AB...
3. Route the other end of the string over the calibration pulley. 4. Attach the free end of the string to the string fixing bolt. 5. When the calibration is complete, return the calibration bracket to the storage position. SITRANS WFS320 Operating Instructions, 02/2016, A5E32880290-AB...
Page 30
• there is no material flow during zero and span calibrations • the test weight is suspended free of any obstructions Accurate calibration is not assured until material tests and a manual span adjustment have been performed, as outlined in the Integrator operating instructions. SITRANS WFS320 Operating Instructions, 02/2016, A5E32880290-AB...
● sensing head outer gasket ● damping fluid ● sensing plate Contact Siemens or your distributor for spare parts ordering information. For a list of parts and part identification diagram, please see Appendix A (Page 45). Unit repair and excluded liability All changes and repairs must be done by qualified personnel and applicable safety regulations must be followed.
If you have any technical questions about the device described in these Operating Instructions and do not find the right answers, you can contact Customer Support: ● Via the Internet using the Support Request: Support request (http://www.siemens.com/automation/support-request) ● Via Phone: –...
Page 34
● Information about field service, repairs, spare parts and lots more under "Services." Additional Support Please contact your local Siemens representative and offices if you have additional questions about the device Find your contact partner at: Local contact person (http://www.siemens.com/automation/partner)
Page 35
Service and maintenance 7.4 Technical support SITRANS WFS320 Operating Instructions, 02/2016, A5E32880290-AB...
3. With the moving beam in the static zero position, thread the range spring until the base just touches the static beam and then turn one complete revolution more. 4. Install the range spring mounting bolt and tighten the range spring locknut. SITRANS WFS320 Operating Instructions, 02/2016, A5E32880290-AB...
To test linearity, at least 3 test weights are used. Each weight represents a different test rate. Record the integrator rate display value associated with each test weight applied to the flowmeter. If all the recorded display values are accurate, the flowmeter measurement is linear. SITRANS WFS320 Operating Instructions, 02/2016, A5E32880290-AB...
Page 38
Some integrators are equipped with a linearization function to compensate for non-linear material flow patterns. Stand-alone linearizing devices are also available for this purpose. Electronic linearization should not be used to correct non-linear test weight results. SITRANS WFS320 Operating Instructions, 02/2016, A5E32880290-AB...
0 to 0.75 V AC from LVDT LVDT conditioner card Output Output 0 to 50 mV DC to Siemens integrator; maximum 300 m (1000 ft) separation between conditioner LVDT conditioner card card and integrator Output 0 - 0.75 V AC @ 2.9 kHz...
ATEX II 3G, Ex nA IIC T6 GcATEX II 2D, Europe proof Ex tb IIIC T70ºC Db IP64 IECEx FMG 13.0016X, Ex nA IIC T6 Gc, International Ex tb IIIC T70 ºC Db IP64 LVDT conditioner card General purpose CE, RCM approvals SITRANS WFS320 Operating Instructions, 02/2016, A5E32880290-AB...
Support plate (customer supplied) Note 1. The sensing head support plate should be rigid and independent of the flowmeter housing. 2. Ensure that the outer gasket seal to the flowmeter housing wall is dust tight. SITRANS WFS320 Operating Instructions, 02/2016, A5E32880290-AB...
Page 45
Dimension drawings 10.1 WFS320 sensing head outline and mounting SITRANS WFS320 Operating Instructions, 02/2016, A5E32880290-AB...