This publication details replacement procedures for
the following AZ40 orifice and seals kit (AZ400 752).
Before carrying out any procedures, see Section 3.
The following procedures must be carried out by a
suitably-trained technician:
• Cleaning/Replacing CO and O2 orifices:
Sections 4, 5, 6 and 7 (pages 4, 4, 6 and 7)
• Replacing sensor assembly gaskets:
Sections 4, 5, 6 and 8 (pages 4, 4, 6 and 8)
• Cleaning / Replacing sensor assembly aspirator:
Sections 4, 5, 6 and 8.4 (pages 4, 4, 6 and 12)
• Replacing sensor assembly flow disk:
Sections 4, 5, 6 and 8.5 (pages 4, 4, 6 and 14)
Kit contains:
• Orifices, gaskets, seals
• This publication
Tools required
• Transmitter door key (supplied with transmitter)
• Operating instruction OI/AZ40-EN*
• 3/16 in, 1/4 in, 5/32 in allen keys (wrenches)
• 7/8 in open-ended spanner (wrench)
• 10 in adjustable spanner (wrench)
• Medium flat-bladed screwdriver
• 2 Small screwdrivers
(1 × flat-bladed, 1 × crosshead)
• Single-edged razor blade
• Anti-seize compound
(suitable for temperatures up to 200 °C [392 °F])