3.7 Combined operation
NOTICE
Installation of analyzer modules
Analyzer modules that are installed in a device version other than the one for which they were
intended can cause damage. Therefore:
● Prior to installation, check the device version for which the analyzer module is intended.
● Only install analyzer modules manufactured for operation in a rack-mounted device into
● Only install analyzer modules manufactured for operation in a wall-mounted device into
Additional information is also available from your Siemens partner → Technical support
(Page 144).
Slots for analyzer modules
The figure below shows the slots for the analyzer modules:
R
W
Figure 3-10
The figure below shows the slots for the analyzer modules:
40
You will find the corresponding information in the MLFB number on the module nameplate:
– 7MB3010-xxxxx-xxAx-Z → "A": Analyzer module for rack-mounted devices
– 7MB3010-xxxxx-xxBx-Z → "B": Analyzer module for wall-mounted devices
such a device.
such a device.
Rack-mounted device (rear)
Wall-mounted device (bottom)
Slot 1 (AM1)
Slot 2 (AM2)
Slots for analyzer modules
Rack-mounted device
Operating Instructions, 05/2015, A5E31874006-03