To prevent your m-PIMA™ Analyser from being
damaged due to a sudden and uncontrolled
power cut a controlled shut down process
will be initiated automatically which puts the
internal battery into its EP-mode when a critical
low limit is reached (please refer to page 13 to
learn how to deactivate the EP-mode).
Note: Independent from current processes the
m-PIMA™ Analyser will stop all activities
and turn itself off, even during an
ongoing test run!
• Remove cartridge and connect your
m-PIMA™ Analyser to A/C power to
completely recharge the internal battery
before you continue to work with the
m-PIMA™ Analyser.

Export Errors

If the «EXPORT FAILED» screen appears, an
unspecified error has occurred during the
export of results. This may have been caused
by either the storage device or the m-PIMA™
Analyser.
• Press
Should this fail a second time insert a
different USB storage device and export
results.
Should this error appear when using a
10
Mobile Modem press
export of results.
Please also refer to the respective product information for further details and
options.
68
i
and repeat the export of results.
i
and repeat the
2019-11-07
WARNING
Controlled shutdown initiated.
2019-11-07
ERROR
EXPORT FAILED
10 WARNINGS & ERROR MESSAGES