Multimobil 10 Installation Instructions Version: 4.0 The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable damages. Siemens Ltd. Med Division India Version 4.0 Page 1 of 12...
SWITCHING THE UNIT ON ......................7 :.................... 8 ORMING OF THE APACITOR :..........................9 OLLIMATOR :........................... 10 OVEMENTS ..........................10 XPOSURE 2.4.1 Sample Exposure: ......................11 HANDING OVER MULTIMOBIL 10.................... 12 Siemens Ltd. Med Division India Version 4.0 Page 2 of 12...
The radiation indication lamp should light only when the exposure is released. In case the lamp lights in any other condition, the unit should be immediately switched off and SIEMENS SERVICE DEPARTMENT should be informed. Any change / or replacement in the unit must be carried out by the manufacturer or a person authorised by manufacturer ONLY.
(except handles) indicates possible danger of upward force the order of 25kg, when the arm is released from its locked position. In such a situation, it is recommended to suspend the installation and contact FACTORY for further clarification. Siemens Ltd. Med Division India Version 4.0 Page 4 of 12...
Bleeder resistor to X9.D950A 3 pin connector discharge cable The max. Permissible resistance between the earthing terminal in the control and the single tank is less than 0.2 ohms. Siemens Ltd. Med Division India Version 4.0 Page 6 of 12...
After set-up, initial hardware checks and capacitor bank fully charged to 355V DC the display changes to the default kV and mAs settings. The kV & mAs display and settings will now respond to the corresponding switches on the panel. Siemens Ltd. Med Division India Version 4.0 Page 7 of 12...
ST link (jumper pin) on D915 PCB. Switch ON the MCB Switch the Generator ON by actuating momentarily Then the kV and mAs display read as shown in picture, if there are no initialisation Errors. Siemens Ltd. Med Division India Version 4.0 Page 8 of 12...
Open collimator flaps to the maximum. The light will show square, sharp edges and clear cross wires. Close the flaps fully. The light field will show a small square. Siemens Ltd. Med Division India Version 4.0 Page 9 of 12...
: V23 Error indication : V24 In case of an error the corresponding code will be displayed on the seven-segment display. Refer the service instructions to rectify the error. Siemens Ltd. Med Division India Version 4.0 Page 10 of 12...
TP :kV and JR. Also note the T exp from kV waveform. Following should be the values. Kvist : 1v=20kV Texp = 100ms JR : (1V = 50mA) 3.2V Texp = 100ms Siemens Ltd. Med Division India Version 4.0 Page 11 of 12...