Cellular Antenna Quick Start Guide
HP 8712ET/ES and HP 8714ET/ES
This quick start guide provides basic instructions on how to verify the performance of cellular antenna
systems. Refer to the Option 100 Fault Location and Structural Return Loss Measurement User's Guide
Supplement for more detailed information. Also, please refer to your analyzer's User's Guide for safety,
warranty, and assistance information.
NOTE
This quick start guide assumes the use of an HP 8712ET or HP 8714ET. If you are using an HP 8712ES or
HP 8714ES, some key presses and displays will be slightly different.
To Make a Fault Location Measurement
1. Choose the measurement parameters or recall an instrument state.
PRESET
BEGIN
Start Distance
XX
2. Reduce the interference.
Cable
Fault Location
BEGIN
Band Pass Center Frequency
MHz
Fault Loc Frequency
FREQ
Band Pass Max Span
AVG
System Bandwidth
Narrow 250 Hz
HP Part Number 08714-90022
Printed in USA October 1998
© Copyright 1998 Hewlett-Packard Company
Cable
Fault Location
Stop Distance
ENTER
900
MHz
300
XX
ENTER
3. Calibrate the analyzer.
CAL
-1-
Full Band Cal

Advertisement

loading

Summary of Contents for HP 8714ET

  • Page 1 NOTE This quick start guide assumes the use of an HP 8712ET or HP 8714ET. If you are using an HP 8712ES or HP 8714ES, some key presses and displays will be slightly different. To Make a Fault Location Measurement 1.
  • Page 2 4. Connect the equipment. 5. Interpret the measurement / save results. Marker Search Max Search MARKER Search − > Max Define Save Data ON SAVE RECALL Prior Menu Save State If Measurement Results Were Poor The following results indicate a specific problem you may encounter. Refer to your Option 100 User’s Guide Supplement for more information.
  • Page 3 To Make a Return Loss Measurement 1. Choose the measurement parameters or recall an instrument state. PRESET MEAS 1 Reflection System Bandwidth Narrow 250 Hz 2. Calibrate the analyzer. 3. Connect the equipment. 1-Port User 1-Port 4. Interpret the measurement / save results. 5.
  • Page 4 To Make an Insertion Loss Measurement 1. Choose the measurement parameters or recall an instrument state. Reflection PRESET MEAS 1 System Bandwidth Narrow 250 Hz 2. Calibrate the analyzer. 3. Connect the equipment. 1-Port User 1-Port 4. Interpret the measurement / save results. 5.

This manual is also suitable for:

8712es8712et8714es