6.2.8 Single point temperature and pressure calibration

Calibrating temperature

l
Calibrating pressure
l
6.2.8.1 Calibrating temperature
To calibrate temperature:
Unlock the keypad and type the PASSCODE to enter level 3 mode. Refer to the section "
l
level 3 mode
" for accessing level 3 mode.
By default, the following appears on the display.
l
MAIN MENU
ALARMS
Using the UP arrow and DOWN arrow scroll through the options in level 3 mode until CALIB P/T
l
appears.
Press OK. The following appears on the display.
l
CALIB P/T
CALIB PRES
Press the DOWN arrow, the following appears on the display.
l
CALIB P/T
CALIB TEMP
Press OK, the temperature read by EC350 appears on the display.
For example:
75.20
F
LIVE TEMP
Here, 75.20 is the value that EC350 reads and F stands for the unit of temperature that is
Fahrenheit.
Press OK. The following appears on the display.
l
75.20
F
INPUT REFT