hit counter script

Husqvarna K 760 Workshop Manual page 25

Hide thumbs Also See for K 760:
Table of Contents

Advertisement

10
A
25
CARBURETTOR
Functional test
Pressure tester 501 56 27-01
The pressure tester has a pump piston that is operated with one hand.
The pressure tester is supplied with an adapter nipple for small sizes.
Make the hose connection to the nipple as short as possible, this gives a
clearer test result.
Needle valve
Reliable inspection of the needle valve's function demands certain conditions.
The clearest result is obtained from a carburettor that is drained of fuel, yet
still has some fuel residue. A completely "dry" carburettor, or a carburettor
that has not been used for a long time give a measurement result that is
difficult to interpret.
Tip when the carburettor is full with fuel: Turn the carburettor with the
pulse line upwards, as shown in the picture. Pump up the pressure to approxi-
mately 30 psi/200 kPa and let the fuel run out through the venturi under
water. Not until air bubbles emerge from the venturi, can the function of the
needle valve be tested as set out below.
Water bath
By lowering the carburettor into the water any external leakage can be
examined at the same time as the function of the needle valve. Connect a
piece of hose to the pulse line, this should come out above the water line.
Connect a thin, short hose to the carburettor's fuel pipe. Pump a pressure
to 15–36 psi/100–250 kPa, the needle valve should then lift. Air bubbles then
come out of the jets in the venturi. The pressure should then drop to approxi-
mately 15–7 psi/100–50 kPa and then drop significantly slower or stop com-
pletely. Pressure drop after 7 psi/50 kPa indicates a leaking needle valve.
If the needle valve does not open at 36 psi/250 kPa (max. permitted pres -
sure test) the needle valve has jammed. This can usually be forced open by
blowing gently through the hole for filter compensation (A).
If no pressure can be pumped up, this may be due to a punctured pump
diaphragm.
Leaking needle valve – indications
A leaking needle valve can be seen in numerous ways. A machine that is run
for short periods is often difficult to start when it is warm, fuel leaks into the
venturi and gives too much fuel for the next warm start. With a cold start,
especially when the machine has not been used for a long period, the fuel
chamber above the needle has been slowly emptied and fuel has evaporated.
First after several attempts to start, has new fuel been pumped forwards and
the machine starts.
Pump diaphragm
Connect the pressure tester to the fuel pipe. Insert a hose from the impulse
line and pump from the fuel connection. If air comes out of the impulse
channel the pump diaphragm is leaking.
Measurement chamber diaphragm
The condition of the diaphragm cannot be checked by pressure measurement.
A punctured diaphragm stops the fuel supply.
Air purge – non-return valve
A valve in the carburettor prevents fuel from leaking out of the carburettor if
leaks should occur in the air purge or connections. Place a hose on the carbu-
rettor's air purge connection, suck and blow. The valve must open at low
underpressure and close at overpressure.

Advertisement

Table of Contents
loading

This manual is also suitable for:

K 760 cut-n-break

Table of Contents

Save PDF