Table of Contents ILE-61 Sensing Head ......................1 Specifications ........................2 Installation ........................... 4 Sensing Head ............................4 Sensing Plate............................4 Viscous Damper............................5 Interconnection............................5 Non-Hazardous Unit without LVDT Conditioner Card ............5 Non-Hazardous Unit with Sensing Head Mounted LVDT Conditioner Card ....6 Non-Hazardous and Hazardous Units with Remote-Located LVDT Conditioner Card 7 Calibration ..........................
ILE-61 Sensing Head The Milltronics ILE-61 sensing head is an out-of-process sensing element for A, E, and V Series solids flowmeters. It is used for continuous in-line weighing of powdered or granular dry bulk solid materials. The material is directed toward the sensing plate. The horizontal impact force of the material, deflecting the sensing plate, displaces the core of the sensing head LVDT (linear variable differential transformer).
Epoxy paint, synergistic polymer, or PFA coating of external aluminum casting surfaces • Epoxy painted mild steel or stainless steel rear cover Approvals • • Optional CSA. Class I, Groups C and D; Class II, Groups E, F, and G Page 2 Milltronics ILE-61 – INSTRUCTION MANUAL 7ML19985CX01...
Page 7
UHMW polyurethane or Ceramic Tile abrasion resistant lining • Synergistic polymer or PFA coating for low cohesion and friction LVDT Conditioner Card Power • ± 5 V DC (typically from Siemens Milltronics integrator) Ambient Temperature • –40 to 50 °C (–40 to 122 °F) Input •...
Insert the taper pin (flat side down), from the left side. Tighten the right taper pin nut to lock the sensing plate shaft in place. Tighten the left taper pin nut. shaft socket flat side up taper pin (flat side down) Page 4 Milltronics ILE-61 – INSTRUCTION MANUAL 7ML19985CX01...
Run additional conductors: - from SF 500 terminal 12 to conditioner terminal block marked“Integrator +EXC” - from SF 500 terminal 13 to conditioner terminal block marked “Integrator –EXC” For further connection information on specific LVDTs, consult Siemens Milltronics. ♦ Note: Shields are common, but not grounded to chassis.
Run additional conductors: - from SF 500 terminal 12 to conditioner terminal block marked “Integrator +EXC” - from SF 500 terminal 13 to conditioner terminal block marked “Integrator –EXC” For further connection information on specific LVDTs, consult Siemens Milltronics. ♦ Note: Shields are common, but not grounded to chassis.
If this procedure is performed after the integrator is calibrated, a new integrator zero and span calibration, span adjust, and Factoring (if required) should be performed. sensing plate changes < 0.01 V AC test weight 7ML19985CX01 Milltronics ILE-61 – INSTRUCTION MANUAL Page 9...
When the calibration is complete, return the calibration bracket to the storage position. string fixing bolt calibration arm test weight Note: Accurate calibration is not assured until material tests and a span adjust are performed. Page 10 Milltronics ILE-61 – INSTRUCTION MANUAL 7ML19985CX01...
Please note the following: • The user is responsible for all changes and repairs made to the device. • All new components must be provided by Siemens Milltronics Process Instruments Inc. • Restrict repair to faulty components only.
10. Install the new inner gasket and reverse the procedure (steps 1 through 9). 11. Perform the LVDT output zero procedure. 12. Referring to the integrator instruction manual, perform an integrator zero and span calibration. Perform a span adjust if calibration accuracy appears affected. Page 12 Milltronics ILE-61 – INSTRUCTION MANUAL 7ML19985CX01...
Perform the LVDT output zero procedure. Refer to Calibration. Perform an integrator zero and span calibration. Refer to the integrator manual, Calibration. Perform a span adjust and factoring as required. Refer to the integrator manual, Calibration. 7ML19985CX01 Milltronics ILE-61 – INSTRUCTION MANUAL Page 13...
Leaf springs are damaged Replace leaf springs, recalibrate flowmeter, and integrator Material flow patterns vary Consult Siemens Milltronics or your distributor Accuracy varies with mate- Non-linear operation Refer to Linearity rial flowrate Page 14 Milltronics ILE-61 – INSTRUCTION MANUAL 7ML19985CX01...
Some integrators are equipped with a linearization function to compensate for non-linear material flow patterns. Stand alone linearizing devices are also available for this purpose. Note: Electronic linearization should not be used to correct non-linear test weight results. 7ML19985CX01 Milltronics ILE-61 – INSTRUCTION MANUAL Page 15...