Reference operating conditions according to IEC 60770-1
Ambient temperature
Humidity
Ambient pressure
11.5
Process
Temperature and pressure
Note
Pressure
Universal stamped flanges are to be used for 0.5 bar max pressure only.
Version
40 m
100 m
Maximum and minimum temperatures are dependent on the process connection, antenna
1)
and O-ring materials. Use of the Easy Aimer limits maximum temperature.
11.6
Communication: Profibus PA
Communication: Profibus PA 3.01
LR560 (PROFIBUS PA)
Operating Instructions, 06/2018, A5E34648471-AE
1)
Stainless steel flange
–40 to +100 °C
(–40 to +212 °F)
–40 to +200 °C
(–40 to +392 °F)
15 to 25 °C (59 to 77 °F)
45 to 75% relative humidity
860 to 1060 mbar a (86000 to 106000 N/m
Aimer flange
0.5 bar max.
–40 to +100 °C
(–40 to +212 °F)
–40 to +200 °C
(–40 to +392 °F)
Technical data
11.5 Process
g)
2
Aimer flange
3.0 bar max
–40 to +100 °C
(–40 to +212 °F)
–40 to +120 °C
(–40 to +248 °F)
151