Page 1
Introduction Safety notes Installing/mounting SITRANS Connecting Temperaturmessumformer SITRANS TH320/TH420/TR320/ Commissioning TR420/TF320/TF420 Service and maintenance Compact Operating Instructions Technical specifications Appendix A 03/2018 A5E41865021-AA...
Page 2
Note the following: WARNING Siemens products may only be used for the applications described in the catalog and in the relevant technical documentation. If products and components from other manufacturers are used, these must be recommended or approved by Siemens. Proper transport, storage, installation, assembly, commissioning, operation and maintenance are required to ensure that the products operate safely and without any problems.
Table of contents Introduction..............................7 Purpose of this documentation....................7 Document history........................7 Scope of validity of this document....................7 Product compatibility........................8 Intended use..........................8 Nameplate layout........................10 1.6.1 Nameplate layout SITRANS TH.....................10 1.6.2 Nameplate layout SITRANS TR.....................11 1.6.3 Nameplate layout SITRANS TF.....................12 Checking the consignment.....................12 Security information.......................13 Transportation and storage....................13 1.10...
Page 4
Table of contents 3.6.1 Mounting device to a pole with mounting bracket..............26 3.6.2 Mounting device to the wall with mounting bracket..............27 Disassembly...........................29 Connecting..............................31 Basic safety notes........................31 4.1.1 Connecting device in energized state in hazardous areas.............32 Connecting TH320.........................33 Connecting TH420.........................34 Connecting TR320.........................35 Connecting TR420.........................37 Connecting the SITRANS TF dual chamber housing.............39 4.6.1...
Page 5
Table of contents Construction...........................65 Common specifications......................66 Certificates and approvals......................67 7.4.1 ATEX/IECEx and related......................67 7.4.2 FM/CSA..........................72 Display...........................75 Appendix A..............................77 Technical support........................77 Certificates..........................77 Index................................79 SITRANS TH320/TH420/TR320/TR420/TF320/TF420 Compact Operating Instructions, 03/2018, A5E41865021-AA...
Introduction Purpose of this documentation These instructions are a brief summary of important features, functions and safety information, and contain all information required for safe use of the device. Read the instructions carefully prior to installation and commissioning. In order to use the device correctly, first review its principle of operation.
Introduction 1.5 Intended use Product compatibility The following table describes the compatibility between the edition of this manual, the device revision, the engineering system and the associated EDD. Manual Comments Device revision Compatible version of device integration package edition 03/2018 First edition HART 7 SIMATIC PDM V9.1...
Page 9
Introduction 1.5 Intended use The output signal is a load-independent direct current from 4 to 20 mA in accordance with the input characteristic, superimposed by the digital HART signal. The dual input mode also supports drift detection of the inputs, whereby maintenance intervals can be more easily planned.
Introduction 1.6 Nameplate layout Nameplate layout 1.6.1 Nameplate layout SITRANS TH ① Conformity with country-specific guidelines ② Manufacturer's address ③ Place of manufacture ④ Read the compact operating instructions ⑤ CE marking with ID number of the named testing authority ⑥...
Introduction 1.6 Nameplate layout 1.6.2 Nameplate layout SITRANS TR ① ⑦ Manufacturer CE marking with ID number of the named testing authority ② ⑧ QR code for mobile website with device-spe‐ Device version cific information about the product ③ ⑨ Conformity with country-specific guidelines Serial number ④...
Nameplate layout SITRANS TF TEMPERATURE TRANSMITTER (S) N1J7002729 SITRANS TF320 HART SIL (1P) 7NG0320-0BA00-0AA0-Z DEVICE NO: nn.nn.nn. Siemens AG C11+C20+D10+D81+D90+P10+V01+W02+Y04 DE-76181 Karlsruhe Made in France UB: DC 13.1...32V; lo: 4...20mA+HART IP66/68 II 1 G Ex ia II C T6/ T5/T4 Ga or...
Siemens’ products and solutions undergo continuous development to make them more secure. Siemens strongly recommends that product updates are applied as soon as they are available and that the latest product versions are used. Use of product versions that are no longer supported, and failure to apply the latest updates may increase customer’s exposure to cyber...
The content reflects the technical status at the time of publishing. Siemens reserves the right to make technical changes in the course of further development.
Safety notes Prerequisites for safe use This device left the factory in good working condition. In order to maintain this status and to ensure safe operation of the device, observe these instructions and all the specifications relevant to safety. Observe the information and symbols on the device. Do not remove any information or symbols from the device.
If you need additional information not covered by these instructions, contact your local Siemens office or company representative. Note Operation under special ambient conditions...
Page 17
Safety notes 2.3 Use in hazardous areas WARNING Use in hazardous area Risk of explosion. ● Only use equipment that is approved for use in the intended hazardous area and labelled accordingly. ● Don't use devices that have been operated outside the conditions secified for hazardous areas.
Page 18
Safety notes 2.3 Use in hazardous areas SITRANS TH320/TH420/TR320/TR420/TF320/TF420 Compact Operating Instructions, 03/2018, A5E41865021-AA...
Installing/mounting Basic safety notes 3.1.1 ATEX/IECEx and others 3.1.1.1 Safety information for installation in "Intrinsic safety "ia/ib" Zones 0, 1, 2, 20, 21, 22 and M1 General installation information The first two digits of the serial number stand for the year of production. With aluminum enclosures, you also have to make sure during installation that ignition sources due to sparks caused by impact or friction are prevented.
Installing/mounting 3.1 Basic safety notes Aluminum enclosure are not permitted in mines. The enclosure must be suitable for the application and installed to meet the applicable regulations. Cable entries and dummy elements must meet the same requirements. 3.1.1.2 Safety information for installation in "Non-sparking nA/ec" and "Intrinsic safety ic" Zone 2 and 22 General installation instructions If the enclosure is made of non-metallic materials or of painted metal, electrostatic charging...
Installing/mounting 3.1 Basic safety notes 3.1.2 CSA/FM 3.1.2.1 Safety information for installation in "Intrinsic safety "ia" Hazardous Area CL I, Div 1, GP ABCD CL I, Zone 0 IIC General installation instructions ● Install in accordance with the US the National Electrical Code (NEC) or for Canada the Canadian Electrical Code (CEC).
Installing/mounting 3.1 Basic safety notes 3.1.2.2 Safety information for installation in "Non Incendive nA" Hazardous Area CL I, Div 2, GP ABCD CL I, Zone 2 IIC General installation instructions ● The transmitter must be installed in an enclosure providing a degree of protection of at least IP54 according to IEC 60529 that is suitable for the application and is correctly installed.Cable entry devices and blanking elements shall fulfill the same requirements.
Installing/mounting 3.2 Installing SITRANS TH in a connection head NOTICE Use of line and cable entries made of plastic in hazardous areas Device damage caused by impact at temperatures below -20 °C. ● Make sure that the line and cable entries are protected from impacts. Installing SITRANS TH in a connection head Note The transmitter is only designed for installation in a type B connection head or larger.
Installing/mounting 3.3 Installing SITRANS TH on DIN rail and G rail Note Using the lock washers (SITRANS TH) ④ The lock washers included in the delivery are only required for securely fastening the ② ③ transmitter when the transmitter is directly installed on the round plate of a temperature probe.
Installing/mounting 3.5 Mounting the SITRANS TF single chamber housing Figure 3-1 Securing the transmitter on DIN rails Figure 3-2 Securing the transmitter on G rails Installing SITRANS TR on DIN rail The transmitter is secured to a 35 mm DIN rail in accordance with DIN EN 60715. Comply with the ambient conditions specified in the technical data.
Installing/mounting 3.6 Mounting the SITRANS TF dual chamber housing Procedure ① Mounting holes for mounting to a pipe clamp or wall ② Mounting holes for fastening mounting plate on device Figure 3-3 Mounting device to a pipe clamp or wall with rail 1.
Installing/mounting 3.6 Mounting the SITRANS TF dual chamber housing Procedure ① ⑤ Mounting holes on the enclosure Nut M8 ② ⑥ Mounting holes in the mounting bracket (90° ro‐ Washer DIN 125 - 8.4 tatable) ③ ⑦ Washer DIN 125 - 8.4 Pole clamp for mounting to pole ④...
Page 28
Installing/mounting 3.6 Mounting the SITRANS TF dual chamber housing Procedure ① ④ Mounting holes on the enclosure Fixing screws M8 or 5/16 UNF thread ② ⑤ Mounting holes in the mounting bracket Fixing screws (90° rotatable) ③ Washer DIN 125 - 8.4 Figure 3-5 Mounting device to the wall with mounting bracket ①...
Installing/mounting 3.7 Disassembly Disassembly WARNING Incorrect disassembly The following risks may result from incorrect disassembly: - Injury through electric shock - Risk through emerging media when connected to the process - Risk of explosion in hazardous area In order to disassemble correctly, observe the following: ●...
Connecting Basic safety notes WARNING Unsuitable cables, cable glands and/or plugs Risk of explosion in hazardous areas. ● Use only cable glands/plugs that comply with the requirements for the relevant type of protection. ● Tighten the cable glands in accordance with the torques specified in Technical specifications (Page 65).
Connecting 4.1 Basic safety notes 4.1.1 Connecting device in energized state in hazardous areas WARNING Connecting device in energized state Risk of explosion in hazardous areas. ● Connect devices in hazardous areas only in a de-energized state. Exceptions: ● Devices having the type of protection "Intrinsic safety Ex i" may also be connected in energized state in hazardous areas.
Connecting 4.2 Connecting TH320 Connecting TH320 Procedure 1. Connect the input or inputs to terminals 3 to 6. 2-wire, 3-wire or 4-wire RTD or TC (internal CJC or linear resistance external 2-wire or 3-wire CJC) I1: Input 1 Voltage input 3-wire or 4-wire potentiometer (single-pole or two-pole) 2.
Connecting 4.4 Connecting TR320 Connecting TR320 Procedure 1. Connect the input or inputs to terminals 5 to 8. The terminals are equipped with a coding profile to ensure intrinsic safety. 2-wire, 3-wire or 4-wire RTD or TC (internal CJC or wire resistance external 2-wire or 3-wire CJC) I1: Input 1...
Page 36
Connecting 4.4 Connecting TR320 For devices with HART communication, connect a ≥ 250 Ω resistor. 3. Take the polarity into account. The device has reverse polarity protection. SITRANS TH320/TH420/TR320/TR420/TF320/TF420 Compact Operating Instructions, 03/2018, A5E41865021-AA...
Connecting 4.5 Connecting TR420 Connecting TR420 Procedure 1. Connect the input or inputs to terminals 5 to 12. The terminals are equipped with a coding profile to ensure intrinsic safety. Input 1 (I1) and/or input 2 (I2): Input 1 and/or input 2: Voltage input 2-wire, 3-wire or 4-wire RTD or TC (int.
Connecting 4.6 Connecting the SITRANS TF dual chamber housing 4.6.2 Connecting the SITRANS TF320 dual chamber housing Procedure ① ⑤ Supply voltage Input terminals ② ⑥ HART communication resistor (optional) Protective conductor connector ③ ⑦ Cable gland Ground terminal ④ Test terminals Figure 4-2 Connecting the device...
Page 41
Connecting 4.6 Connecting the SITRANS TF dual chamber housing 5. For devices with "Intrinsic safety" type of protection, use an isolating power supply that corresponds to the requirements of the relevant type of protection. ⑤ 6. Connect the input terminals 2-wire, 3-wire or 4-wire RTD or TC (internal CJC or linear resistance...
Connecting 4.6 Connecting the SITRANS TF dual chamber housing 4.6.3 Connecting the SITRANS TF420 dual chamber housing Procedure ① ⑤ Supply voltage Input terminals ② ⑥ HART communication resistor (optional) Protective conductor connector ③ ⑦ Cable gland Ground terminal ④ Test terminals Figure 4-3 Connecting the device...
Page 43
Connecting 4.6 Connecting the SITRANS TF dual chamber housing 5. For devices with "Intrinsic safety" type of protection, use an isolating power supply that corresponds to the requirements of the relevant type of protection. ⑤ 6. Connect the input terminals Input 1 (I1) and/or Input 1 (I1) and/or Input 1 (I1) and/or...
Connecting 4.7 Connecting the SITRANS TF single chamber housing Connecting the SITRANS TF single chamber housing 4.7.1 Opening the device Procedure ① Cover of the electrical cable compartment ② Safety catch Figure 4-5 Opening device with single chamber housing ② 1.
Connecting 4.7 Connecting the SITRANS TF single chamber housing 4.7.2 Connecting the SITRANS TF320 single chamber housing Procedure 1. Remove the display from the holder. ① 2. Loosen the fastening screws of the temperature transmitter and remove it. ② 3. Lead the shielded connecting cables through the cable glands SITRANS TH320/TH420/TR320/TR420/TF320/TF420 Compact Operating Instructions, 03/2018, A5E41865021-AA...
Page 47
Connecting 4.7 Connecting the SITRANS TF single chamber housing ③ 4. Connect the shield to the screw of the ground terminal The screw of the ground terminal is electrically connected to the external protective conductor connection. 5. Insert the temperature transmitter and screw it in. ④...
Connecting 4.7 Connecting the SITRANS TF single chamber housing 9. Screw the cover back on as far as it will go. ⑦ 10.Connect the device to the plant with the protective conductor connection 4.7.3 Connecting the SITRANS TF420 single chamber housing Procedure 1.
Page 49
Connecting 4.7 Connecting the SITRANS TF single chamber housing ② 3. Lead the shielded connecting cables through the cable glands ③ 4. Connect the shield to the screw of the ground terminal The screw of the ground terminal is electrically connected to the external protective conductor connection.
Page 50
Connecting 4.7 Connecting the SITRANS TF single chamber housing ④ ⑤ 6. Connect the input terminals and output terminals Connecting TH420 (Page 34) ⑥ Figure 4-6 HART communication resistor (optional) 7. For devices with "Intrinsic safety" type of protection, use an isolating power supply that corresponds to the requirements of the relevant type of protection.
Connecting 4.7 Connecting the SITRANS TF single chamber housing 4.7.4 Closing the device Procedure ① Cover ② Safety catch ③ Cable gland Figure 4-7 Closing device with single chamber housing ① 1. Screw the cover back on as far as it will go. ②...
Page 52
Connecting 4.7 Connecting the SITRANS TF single chamber housing SITRANS TH320/TH420/TR320/TR420/TF320/TF420 Compact Operating Instructions, 03/2018, A5E41865021-AA...
Commissioning Basic safety notes WARNING Improper commissioning in hazardous areas Device failure or risk of explosion in hazardous areas. ● Do not commission the device until it has been mounted completely and connected in accordance with the information in Technical specifications (Page 65). ●...
Commissioning 5.3 Commissioning SITRANS TH/TR with HART Procedure 1. If you want to change the factory settings, configure the device using the USB modem and the SIPROM T parameter assignment software. – Read the operating instructions for information on this. 2.
Commissioning 5.4 Commissioning SITRANS TF with 4 ... 20 mA Procedure 1. Install/mount the device. Installing SITRANS TH in a connection head (Page 23) Installing SITRANS TH on DIN rail and G rail (Page 24) Installing SITRANS TR on DIN rail (Page 25) 2.
Commissioning 5.5 Commissioning SITRANS TF with HART 4. Switch on the power supply. Switching on the supply voltage (Page 57) 5. Wait for 2 seconds. After this start-up time, the device is operational. 6. Wait 5 minutes to get exact measured values. Commissioning SITRANS TF with HART Introduction In this section, you will learn how to commission the device step-by-step.
Commissioning 5.6 Switching on the supply voltage Switching on the supply voltage Requirement ● You have connected the device correctly. (Page 31) ● The terminal voltage on the device is correct. (Page 65) Procedure Switch on the power supply. ● Product name and firmware version appear briefly on the display. ●...
Page 58
Commissioning 5.6 Switching on the supply voltage SITRANS TH320/TH420/TR320/TR420/TF320/TF420 Compact Operating Instructions, 03/2018, A5E41865021-AA...
Service and maintenance Basic safety notes 6.1.1 Maintenance The device is maintenance-free. However, a periodic inspection according to pertinent directives and regulations must be carried out. An inspection can include check of: ● Ambient conditions ● Seal integrity of the process connections, cable entries, and cover screws ●...
Service and maintenance 6.2 Cleaning CAUTION Releasing button lock Improper modification of parameters could influence process safety. ● Make sure that only authorized personnel may cancel the button locking of devices for safety-related applications. NOTICE Penetration of moisture into the device Device damage.
6.3.1 SITRANS TF WARNING Impermissible repair and maintenance of the device ● Repair and maintenance must be carried out by Siemens authorized personnel only. WARNING Impermissible repair of explosion protected devices Risk of explosion in hazardous areas ● Repair must be carried out by Siemens authorized personnel only.
Incorrect measured values will be displayed. Changing the seals in a process flange of a differential pressure measuring cell can alter the start-of-scale value. ● Changing seals in devices with differential pressure measuring cells may only be carried out by personnel authorized by Siemens. Note Using the wrong seals Using the wrong seals with flush-mounted process connections can cause measuring errors and/or damage the diaphragm.
Enclose the bill of lading, return document and decontamination certificate in a clear plastic pouch and attach it firmly to the outside of the packaging. Required forms ● Delivery note ● Return document (http://www.siemens.com/processinstrumentation/returngoodsnote) with the following information: – Product (item description) – Number of returned devices/replacement parts –...
Page 64
Service and maintenance 6.5 Disposal SITRANS TH320/TH420/TR320/TR420/TF320/TF420 Compact Operating Instructions, 03/2018, A5E41865021-AA...
Technical specifications 7.3 Common specifications Common specifications General specifications SITRANS TH/TR SITRANS TF Supply voltage ... 48 V DC Single chamber housing Dual chamber housing ... 30 V DC (Ex i) Without display: Without display: 7.5 ... 48 V DC 8.2 ...
Technical specifications 7.4 Certificates and approvals Certificates and approvals 7.4.1 ATEX/IECEx and related ATEX/IECEx and others SITRANS TH/TR SITRANS TF Display Certificate number EN 60079-0:2012, DEKRA 17 ATEX 0116 X DEKRA xx ATEX xxxx X DEKRA xx ATEX xxxx U A11:2013, EN60079-11:2012 EN60079-15:2010,...
Page 69
Technical specifications 7.4 Certificates and approvals ATEX/IECEx and others SITRANS TH/TR SITRANS TF Display Input terminals = 7.2 V DC, I = 12.9 mA, P = 23.3 mW, L = 200 mH, C 13.5 μF SITRANS TH/TF: 3, 4, 5, 6, 7, 8, 9 SITRANS TR: 5, 6, 7, 8, 9, 10, 11, 12 Zones 2 and 22...
Page 70
Technical specifications 7.4 Certificates and approvals ATEX/IECEx and others SITRANS TH/TR SITRANS TF Display Zones 2 and 22 ● "Non-sparking nA/ec" type of protection Take measures to ensure that the supply voltage (including transients) does not rise above 140% of the rated voltage. Name II 2 G Ex nA IIC T6 ...
Page 71
Technical specifications 7.4 Certificates and approvals ATEX/IECEx and others SITRANS TH/TR SITRANS TF Display Name II 2 G Ex db IIC T6 ... T4 Gb II 2 D Ex tb IIIC T100 °C Db ● ATEX Ex db IIC T6 ... T4 Gb ●...
Page 73
Technical specifications 7.4 Certificates and approvals FM/CSA SITRANS TH/TR SITRANS TF Display Name IS,CL I, Div 1, GP ABCD, T6 ... T4 Ex ia IIC T6 … T4 Ga AEx ia IIC T6 … T4 Ga Ex ib [ia Ga] IIC T6 … T4 Gb AEx ib [ia Ga] IIC T6 …...
Page 74
Technical specifications 7.4 Certificates and approvals FM/CSA SITRANS TH/TR SITRANS TF Display Output terminals = 30 V DC, L = 0 μH, C 1.0 nF SITRANS TH/SI‐ TRANS TF single chamber housing: 1, 2 SITRANS TR: 3, 4 SITRANS TF dual chamber housing: (+) and (-) -50 …...
Technical specifications 7.5 Display FM/CSA SITRANS TH/TR SITRANS TF Display -50 … +85 °C (-58 … +185 °F) ● Permissible temperature class T4 ambient -50 … +75 °C (-58 … +167 °F) temperature temperature class T5 -50 … +60 °C (-58 … +140 °F) temperature class T6 Input terminals = 7.2 V DC...
Page 76
Technical specifications 7.5 Display Display Decimal point change Automatically Error warning According to NAMUR NE 107 Electromagnetic compatibility According to DIN EN 61326-1 and NAMUR recom‐ mendation NE21 Construction Dimensions Diameter: 64.3 mm (2.53") Height: 15 ... 22.5 mm (0.6 ... 0.88") Installation Plug in Enclosure material...
In addition to our documentation, Siemens provides a comprehensive support solution at: ● Service&Support (http://www.siemens.com/automation/service&support) Personal contact If you have additional questions about the device, please contact your Siemens personal contact at: ● Partner (http://www.automation.siemens.com/partner) To find the personal contact for your product, go to "All Products and Branches" and select "Products &...