1.2 Usable Cable Lengths
Four configurations of cable lengths are available for order. Measure the distances between the
equipment and compare them to the distances specified in the tables. If the your distance is less
than the short distance specified, the Site Option is "Short"; otherwise, the Site Option is "Long."
Table 6-3: Order Configuration Options
Configuration
Table 6-4: Usable Cable Lengths
Point A
CRY
Magnet Monitor
ISC, top panel
ISC, top panel
ICC, top panel
ISC, top panel
ISC, top panel
ICC, Pen Panel
ISC Pen Panel
ICC Pen Panel
ICC Pen Panel
MRU
130
Equipment Room — Site Option
A
B
C
D
Point B
Magnet Monitor
ICC
Magnet Monitor
GOC, rear panel
GOC, rear panel
ICC, top panel
Magnet room door switch
E-off switch, Control room or
Equipment room
MAG, gradient cable clamp
block
MAG
E-off switch, Magnet Room
MAG, gradient cable clamp
block
Short
Long
Short
Long
Site Option: Short
Equipment Room
15.7 (400)
Magnet Room
210.6 (5400)
1 MR System Interconnects Specifications
SIGNA Voyager Pre-Installation
Direction 5680008–1EN, Revision 2
Magnet Room — Site Option
Short
Short
Long
Long
Site Option: Long
in. (mm)
370.1 (9400)
661.4 (16800)
645.7 (16400)
97.5 (25000)
97.5 (25000)
252 (6400)
960.6 (24400)
661.4 (16800)
366.6 (9400)
511.8 (13000)
1173.2 (29800)
972.4 (24700)