Technical specifications
Ambient temperature
Ambient pressure
Sample gas flow
Sample gas humidity
Site of installation
Reference gas
Reference application
For devices with filled reference gas side
1)
The technical specifications for time and measuring response as well as for the influencing variables can sometimes differ
2)
significantly for other gas mixtures
13.2
Technical specifications of rack-mounted housing
Table 13-1
Rack-mounted housing: General technical specifications
General
Installation position
Weight
Degree of protection
Electrical characteristics
Auxiliary power supply
Power consumption
EMC interference im‐
munity (electromag‐
netic compatibility)
during operation:
Electrical safety
Rack-mounted device
Operating Instructions, A5E31874006-12
OXYMAT 7
Nitrogen (N
-
Horizontal
Approx. 8.5 kg
IP20 according to EN 60529
with one and/or two OXY‐
MAT 7 / ULTRAMAT 7 units
with one or two CALOMAT 7
units
ULTRAMAT 7
25 °C (77 °F)
Atmospheric (approx. 1 000 hPa)
0.6 Nl/min; Nl = Standard liter
Dew point < -40 °C (-40 °F)
Free of vibration, shock and impact
)
Standard: Nitrogen (N
2
Option: freely selectable
100 to 240 V AC (nominal range of use 85 to 264 V)
50 to 60 Hz (nominal range of use 47 to 63 Hz)
Max. 280 VA
in accordance with the standard requirements of NA‐
MUR NE21 (05/2006) and EN 61326-1 (2013)
in accordance with standard requirements EN 61326-1 (2013)
in accordance with EN 61010-1, overvoltage category II
CALOMAT 7
)
2
1)
-
Hydrogen in nitrogen
13
-
(H
in N
)
2)
2
2
117