Page 2
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho- rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service procedures.
Install or remove assemblies only with the generator switched off and according to ESD guidelines. Preliminary Steps • Set switch (S1) on board D 100 to SS OFF (Fig. 1). Siemens AG RXL2-120.815.01 Page 3 of 6 LITHOSTAR Multiline Medical Solutions Rev.
Page 8
1 - 4 Prerequisites / General Information V36 V35 D110 D160 D100 Fig. 1 LITHOSTAR Multiline RXL2-120.815.01 Page 4 of 6 Siemens AG System manual Rev. 02 02.05 CS PS 24 Medical Solutions...
Page 9
Com.1 BUS mouse Important: - Connect the system controller (D320.XCS) to Com 1. - Connect the PC mouse to the PC BUS mouse connector. Siemens AG RXL2-120.815.01 Page 5 of 6 LITHOSTAR Multiline Medical Solutions Rev. 02 02.05 CS PS 24...
Page 10
D200 D220 D291 D160 F1F2 F3 Com 1 Com 2 Important: The system controller (D200.X5) must be connected to Com 1. LITHOSTAR Multiline RXL2-120.815.01 Page 6 of 6 Siemens AG System manual Rev. 02 02.05 CS PS 24 Medical Solutions...
(phase-to-phase) as follows: • Connect the internal line impedance meter between 2 phases respectively of L1, L2 and • Switch the system breaker ON. • Perform the measurement. Siemens AG RXL2-120.815.01 Page 1 of 4 LITHOSTAR Multiline Medical Solutions Rev. 02 02.05...
Page 12
0.11 Ohm • Switch the system breaker OFF. • Reinstall the knife fuses. • Reinstall the cover over the knife fuses. LITHOSTAR Multiline RXL2-120.815.01 Page 2 of 4 Siemens AG System manual Rev. 02 02.05 CS PS 24 Medical Solutions...
Due to changes in the monitoring of the shock wave system, the NOTE shock wave cabinet must be switched on when switching on the system. The shock wave cabinet can be switched off, when the system is ready for use. Siemens AG RXL2-120.815.01 Page 3 of 4 LITHOSTAR Multiline Medical Solutions Rev.
Page 14
Move the unit to the left so that board D31 is accessible. • Connect control cable X28 on board D31 and secure it with screws. • Reattach the cover. LITHOSTAR Multiline RXL2-120.815.01 Page 4 of 4 Siemens AG System manual Rev. 02 02.05 CS PS 24 Medical Solutions...
Connect the quick couplings and the pump. • Turn on N12 using switch S1; the cooling circuit is filled and ventilated. Ensure that the cooling circuit does not contain air bubbles. NOTE Siemens AG RXL2-120.815.01 Page 1 of 6 LITHOSTAR Multiline Medical Solutions Rev.
Page 16
Leakage in the cooling circuit NOTE ⇒ If the voltage drops Solenoid valve Y1 is defective • Tighten the red cap of the vent valve. LITHOSTAR Multiline RXL2-120.815.01 Page 2 of 6 Siemens AG System manual Rev. 02 02.05 CS PS 24 Medical Solutions...
"Replacing the Sensor Module" of the instructions RXL2-120.091.06.. • Press the coupling bellows together. • Reconnect the elbow. • Set the S2 switch (Fig. 1) to STANDARD MODE (Y3 valve closed). Siemens AG RXL2-120.815.01 Page 3 of 6 LITHOSTAR Multiline Medical Solutions Rev. 02 02.05...
Page 18
Remove air bubbles in the coupling bellows using a syringe (without a needle) (Fig. 3). • Set the S2 switch on board W12 to the EMPTY COUPLING BELLOWS position (Fig. 1) (Y3 valve is open). LITHOSTAR Multiline RXL2-120.815.01 Page 4 of 6 Siemens AG System manual Rev. 02 02.05 CS PS 24 Medical Solutions...
Starting from step 0.1 (default position), increase the shock wave energy in steps up to step 9 and release several shots at each step. • Set the S3 switch back to the STANDARD MODE position. Siemens AG RXL2-120.815.01 Page 5 of 6 LITHOSTAR Multiline Medical Solutions Rev.
Page 20
3 - 6 Shock Wave System This page intentionally left blank. LITHOSTAR Multiline RXL2-120.815.01 Page 6 of 6 Siemens AG System manual Rev. 02 02.05 CS PS 24 Medical Solutions...
• Prior to turning the system over to the customer, save the system programming on the FD - SW POLYDOROS diskette (back-up). Siemens AG RXL2-120.815.01 Page 1 of 8 LITHOSTAR Multiline Medical Solutions Rev.
Remove the jumper from the mAs sockets on board D220 (on the H1 high voltage transformer). • Connect an mAs meter to the test sockets. • Switch the system ON. LITHOSTAR Multiline RXL2-120.815.01 Page 2 of 8 Siemens AG System manual Rev. 02 02.05 CS PS 24 Medical Solutions...
Select "Generator Parameters" in the Adjustment menu. • Select "24" m for the Tube 1 High Tension Cable Length in the Generator Parameters field. • Select "Fluoroscopy" in the "Warm Up" menu. Siemens AG RXL2-120.815.01 Page 3 of 8 LITHOSTAR Multiline Medical Solutions Rev. 02 02.05...
Page 24
• Click OK. • Click "Backup to Disk" in the Data menu. • Insert the Site data floppy. • Click OK. LITHOSTAR Multiline RXL2-120.815.01 Page 4 of 8 Siemens AG System manual Rev. 02 02.05 CS PS 24 Medical Solutions...
, 102 kV, 100 % power at the Operating PC and select 28 mAs. • Select 2V/div for kV and 1V/div for mA on the oscilloscope. • Release an exposure and measure the kV and mA with the oscilloscope. Siemens AG RXL2-120.815.01 Page 5 of 8 LITHOSTAR Multiline Medical Solutions Rev. 02 02.05...
• Remove the universal adjustment device. • Reinsert the rubber plugs unless the shock wave pressure test is being performed. LITHOSTAR Multiline RXL2-120.815.01 Page 6 of 8 Siemens AG System manual Rev. 02 02.05 CS PS 24 Medical Solutions...
Release shock waves. • Remove PTU adapter M and reinsert the rubber plugs. • Move the shock wave head into the park position after completing the measurement. Siemens AG RXL2-120.815.01 Page 7 of 8 LITHOSTAR Multiline Medical Solutions Rev. 02 02.05...
Page 28
4 - 8 Start-Up of Multiline System This page intentionally left blank. LITHOSTAR Multiline RXL2-120.815.01 Page 8 of 8 Siemens AG System manual Rev. 02 02.05 CS PS 24 Medical Solutions...
If it has been assured that the factory settings have not changed, the values in the test certificate and the films that were exposed at the factory can be used for the Acceptance Test Certificate. Siemens AG RXL2-120.815.01 Page 1 of 2...
X and Y can be taken from the Multiline test protocol and used in the Acceptance Test Protocol. LITHOSTAR Multiline RXL2-120.815.01 Page 2 of 2 Siemens AG System manual Rev. 02 02.05 CS PS 24 Medical Solutions...
Select the service software VIDEOMED DIM. → In Menu Adjustment select submenu Gamma calibration Ramp calibration and perform the steps indicated. Siemens AG RXL2-120.815.01 Page 1 of 18 LITHOSTAR Multiline Medical Solutions Rev. 02 02.05 CS PS 24 System manual...
Select IQ start-up and confirm query "DHHS" or "NO" (Menu option § 16 X-Ray Ordinance is presently not used) continue with ESC LITHOSTAR Multiline RXL2-120.815.01 Page 2 of 18 Siemens AG System manual Rev. 02 02.05 CS PS 24 Medical Solutions...
Page 33
RXD-000.038... continue with ESC ADC measurement field Check and enter Light Path. continue with ESC Select Fluoroscopy dose rate Select FL C00 Siemens AG RXL2-120.815.01 Page 3 of 18 LITHOSTAR Multiline Medical Solutions Rev. 02 02.05...
Page 34
Select Videomed DI: Measure the B-Signal Check resolution Protocol Artifacts Corresponding instructions available with the F1 key Help continue with ESC LITHOSTAR Multiline RXL2-120.815.01 Page 4 of 18 Siemens AG System manual Rev. 02 02.05 CS PS 24 Medical Solutions...
Page 35
Sign protocol on page 2, store disk, and protocol in the system binder. • At present, the snap shot for the 33 cm I.I. according to paragraph F can only be recorded in the supplied protocol. Siemens AG RXL2-120.815.01 Page 5 of 18 LITHOSTAR Multiline Medical Solutions Rev.
Confirm the message "IImaximum ... " with OK. • Click Fluoroscopy on. • Change the mA value so that the dose rate desired appears on the dosimeter. LITHOSTAR Multiline RXL2-120.815.01 Page 6 of 18 Siemens AG System manual Rev. 02 02.05 CS PS 24 Medical Solutions...
Page 37
- Call up "Parameter" in the Adjustment menu. - Click Get from Unit. - Select the parameter: am 7 TV-Iris in "Parameter Groups". - Select "TV-Iris format" under Parameter. Siemens AG RXL2-120.815.01 Page 7 of 18 LITHOSTAR Multiline Medical Solutions Rev.
Select Dose rate in the Adjustment menu. • If the message "You are .." appears, confirm with OK. The message "Select II maximum ... " appears. LITHOSTAR Multiline RXL2-120.815.01 Page 8 of 18 Siemens AG System manual Rev. 02 02.05 CS PS 24 Medical Solutions...
Page 39
- Switch on the system. - Record the dose rate again. After each reset of the Multiline, the antiisowatt curve must be NOTE selected on the operating PC. Siemens AG RXL2-120.815.01 Page 9 of 18 LITHOSTAR Multiline Medical Solutions Rev. 02 02.05...
Page 40
End the adjustment, otherwise repeat it. NOTE After each Reset of the Multiline, the antiisowatt curve must be selected on the operating PC. LITHOSTAR Multiline RXL2-120.815.01 Page 10 of 18 Siemens AG System manual Rev. 02 02.05 CS PS 24 Medical Solutions...
If the small chamber is used with the Dali, the correction factor of 0.85 must be included, i.e. measured value x 0.85. • Reinstall the fuse in Z96. • Remove the measurement chamber. Siemens AG RXL2-120.815.01 Page 11 of 18 LITHOSTAR Multiline Medical Solutions Rev. 02 02.05...
Zoom 3 1056 1753 In case of deviations, perform the dose rate setting for DR in accordance with pages 6-13 through 6-16. LITHOSTAR Multiline RXL2-120.815.01 Page 12 of 18 Siemens AG System manual Rev. 02 02.05 CS PS 24 Medical Solutions...
- Repeat this process as many times as necessary to attain the specified PDA size. - Tighten the set screws. - Reset service switch S1 on board D100 at the PDA sensor to the starting position (the service lamp goes out). Siemens AG RXL2-120.815.01 Page 13 of 18 LITHOSTAR Multiline Medical Solutions Rev.
Remove the cover of the light distributor. The room must be darkened for this adjustment, otherwise stray NOTICE light will distort the setting. LITHOSTAR Multiline RXL2-120.815.01 Page 14 of 18 Siemens AG System manual Rev. 02 02.05 CS PS 24 Medical Solutions...
Page 45
With fluoro ON, check whether the value 0 +1 is attained in the field "Difference of expected to actual value". • Otherwise, adjust potentiometer R1 on D100 until the adjustment is successful. Siemens AG RXL2-120.815.01 Page 15 of 18 LITHOSTAR Multiline Medical Solutions Rev.
If case of deviations, change the value of the associated "DR Zoom x" in "am7 TV Iris" under "Parameter Groups" (For larger values, the image becomes brighter and the voltage increases and vice versa.) LITHOSTAR Multiline RXL2-120.815.01 Page 16 of 18 Siemens AG System manual Rev. 02 02.05 CS PS 24 Medical Solutions...
RX63-020.034.08... In addition, steps that must be performed are described in the VIDEOMED DI Service Instructions, RX52-022.061.01... • Switch the system OFF and reattach the cover to the VIDEOMED DI M. Siemens AG RXL2-120.815.01 Page 17 of 18 LITHOSTAR Multiline Medical Solutions Rev.
However, to ensure customer satisfaction, the system must be adapted to the medical applications in question and to the on-site procedures. LITHOSTAR Multiline RXL2-120.815.01 Page 18 of 18 Siemens AG System manual Rev. 02 02.05 CS PS 24 Medical Solutions...
• Perform the start-up for DICOM Connect according to the instructions RXL2-120.814.02. DICOM can be installed in the LITHOSTAR Multiline only if the NOTE MULTISPOT is not installed. Siemens AG RXL2-120.815.01 Page 1 of 2 LITHOSTAR Multiline Medical Solutions Rev. 02 02.05...
• Shockwave release test " test with ECG triggering " test with respiratory triggering " test with respiratory and ECG triggering LITHOSTAR Multiline RXL2-120.815.01 Page 2 of 2 Siemens AG System manual Rev. 02 02.05 CS PS 24 Medical Solutions...
Protective conductor resistance: max. 0.2 Ohm • If it is greater than this value, increase the wire gauge of the protective conductor. • Reduce the transfer resistance (e.g. tighten the screws). Siemens AG RXL2-120.815.01 Page 1 of 2 LITHOSTAR Multiline Medical Solutions Rev. 02 02.05...
Page 52
Test Certificates (unit, generator and IQ Test Certificates). NOTE From serial number 7000 the laser light localizer can be converted to battery operation (see instructions RXL2-120.033.03... ). LITHOSTAR Multiline RXL2-120.815.01 Page 2 of 2 Siemens AG System manual Rev. 02 02.05 CS PS 24 Medical Solutions...
Page 53
Changes to previous version 9 - 1 Chap. 0 Cover sheet, revision, table of contents generated again. Chap. 8 -2 Information regarding installation report included. Siemens AG RXL2-120.815.01 Page 1 of 2 LITHOSTAR Multiline Medical Solutions Rev. 02 02.05 CS PS 24...
Page 54
9 - 2 Changes to previous version This page intentionally left blank. LITHOSTAR Multiline RXL2-120.815.01 Page 2 of 2 Siemens AG System manual Rev. 02 02.05 CS PS 24 Medical Solutions...