Checking the Dose Rate
Siemens AG
Medical Solutions
Select General Data:
Check and add
Customer data
Add Test Info
Fill in Test equipm.
Systemconf. and UNITA have already been
entered correctly at the factory.
continue with ESC
Select Image intensifier
Check and enter Temperature indica-
tor refer to instructions RXD-000.038...
continue with ESC
ADC measurement field
Check and enter Light Path.
continue with ESC
Select Fluoroscopy dose rate
Select FL C00
RXL2-120.815.01
Rev. 02
Page 3 of 18
02.05
CS PS 24
6 - 3
LITHOSTAR Multiline
System manual