1 to 4 1 to 2 1 to 2 1 to 2 1 to 2 1 to 6 1 to 4 1 to 2 MEMOSKOP 2/3/50/100/SUB RX57-029.061.01 Page 2 of 4 Siemens AG Rev. 03 10.97 TD SD 24 Medical Engineering...
• To comply with EMC regulations, PC board shielding must be installed properly. • Always switch the system off or disconnect the power plug prior to removing the covers. Siemens AG RX57-029.061.01 Page 1 of 4 MEMOSKOP 2/3/50/100/SUB Medical Engineering Rev.
Remove the protective conductor and retighten the screw slightly. Rear cover of the SM2000 monitor trolley • Remove the cover screws. • Remove the protective conductor. MEMOSKOP 2/3/50/100/SUB RX57-029.061.01 Page 2 of 4 Siemens AG Rev. 02 10.97 TD SD 34 Medical Engineering...
Realign the contact springs of the basic unit covers with the edges of the cover panels ( SIREMOBIL 2000 only). • Ensure that the protective conductors are properly connected. Siemens AG RX57-029.061.01 Page 3 of 4 MEMOSKOP 2/3/50/100/SUB Medical Engineering Rev.
Perform a functional check and image quality test in accordance with the associated documentation (diagnostic test, IQ binder). Final steps • With the system/unit closed, perform a protective conductor test. Siemens AG RX57-029.061.01 Page 1 of 2 MEMOSKOP 2/3/50/100/SUB Medical Engineering Rev. 02 10.97...
Page 10
2 - 2 Maintenance This page intentionally left blank. MEMOSKOP 2/3/50/100/SUB RX57-029.061.01 Page 2 of 2 Siemens AG Rev. 02 10.97 TD SD 34 Medical Engineering...
LITHOSTAR MULTILINE • Refer to chapter 3 in Operating Service software , RXL2-120.113.01. with UROSKOP D1/D2 • Refer to chapter 11 in Operating Service software, RLL5-310.113.02. Siemens AG RX57-029.061.01 Page 1 of 2 MEMOSKOP 2/3/50/100/SUB Medical Engineering Rev. 02 10.97...
Page 12
3 - 2 Loading the software This page intentionally left blank. MEMOSKOP 2/3/50/100/SUB RX57-029.061.01 Page 2 of 2 Siemens AG Rev. 02 10.97 TD SD 34 Medical Engineering...
You can exit the service menu by pressing the same key combination as for calling up the service menu. LITHOSTAR Multiline • Call up the LITHOSTAR Multiline service menu. Refer to "Setting the software parameters" RXL2-120.032.02... Siemens AG RX57-029.061.01 Page 1 of 2 MEMOSKOP 2/3/50/100/SUB Medical Engineering Rev. 03 10.97...
Press keys <CTRL> + <E>. Exit set-up mode • Switch the system "OFF" and back "ON" again. The memory returns to normal operating mode. MEMOSKOP 2/3/50/100/SUB RX57-029.061.01 Page 2 of 2 Siemens AG Rev. 03 10.97 TD SD 34 Medical Engineering...
Enter the numbers without any blanks. • Press the keys <CTRL> + <E>. Exit set-up mode • Switch the system "OFF" and "ON" again. The memory returns to normal operating mode. Siemens AG RX57-029.061.01 Page 1 of 2 MEMOSKOP 2/3/50/100/SUB Medical Engineering Rev. 02 10.97...
5 - 2 Electrical settings/programming LITHOSTAR Multiline • For the configuration of Memoskop 3 / (100) refer to "Setting the software parameters" RXL2-120.032.02... MEMOSKOP 2/3/50/100/SUB RX57-029.061.01 Page 2 of 2 Siemens AG Rev. 02 10.97 TD SD 34 Medical Engineering...
The customer data is now stored automatically in the Memoskop. • For SIREMOBIL 2000: - Perform Download according to chapter 6 of the instructions RR2-120.061.02.. . - Perform Configuration according to chapter 5 of the instructions RR2-120.061.02... Siemens AG RX57-029.061.01 Page 1 of 6 MEMOSKOP 2/3/50/100/SUB Medical Engineering Rev.
• Remove board D21, insert the new board and reattach the tilt levers. • No service work is necessary. • Close the MEMOSKOP. MEMOSKOP 2/3/50/100/SUB RX57-029.061.01 Page 2 of 6 Siemens AG Rev. 03 10.97 TD SD 34 Medical Engineering...
Attach the new D23 board. • Reinstall all plug connections. • Reinsert D22 without damaging the flat ribbon cable. • Close the MEMOSKOP. Siemens AG RX57-029.061.01 Page 3 of 6 MEMOSKOP 2/3/50/100/SUB Medical Engineering Rev. 03 10.97 TD SD 34...
• Insert the new switching control. • No additional adjustments are necessary. • Reattach the cover with the screws. • Close the MEMOSKOP. MEMOSKOP 2/3/50/100/SUB RX57-029.061.01 Page 4 of 6 Siemens AG Rev. 03 10.97 TD SD 34 Medical Engineering...
32-34 jumpered open All other existing jumpers remain open If applicable Substitute jumper, if applicable Software VA07A must be loaded in the MEMOSKOP. Siemens AG RX57-029.061.01 Page 5 of 6 MEMOSKOP 2/3/50/100/SUB Medical Engineering Rev. 03 10.97 TD SD 34...
Replace board D18 according to chapter 6 of the instructions RR2-120.061.02. • On UROSKOP D1 - Replace board D18. - Perform Download according to the instructions RLL5-310.113.02. MEMOSKOP 2/3/50/100/SUB RX57-029.061.01 Page 6 of 6 Siemens AG Rev. 03 10.97 TD SD 34 Medical Engineering...
Then press the [CTRL] + [E] keys simultaneously. • Then, depending on the requirement, program the contact time TK and/or postcontact time TN or exit the Hardcopy Setup (refer to the section "Exit Setup "). Siemens AG RX57-029.061.01 Page 1 of 4 MEMOSKOP 2/3/50/100/SUB Medical Engineering Rev.
After switching the unit back on, the new Hardcopy timing will be valid. • If plug D1.X6 was removed from the MEMOSKOP, reconnect it. MEMOSKOP 2/3/50/100/SUB RX57-029.061.01 Page 2 of 4 Siemens AG Rev. 02 10.97 TD SD 24 Medical Engineering...
Page 28
8 - 2 Changes as compared to previous versions This page intentionally left blank. MEMOSKOP 2/3/50/100/SUB RX57-029.061.01 Page 2 of 2 Siemens AG Rev. 02 10.97 TD SD 34 Medical Engineering...