Functional checks............19 5 _______ Changes since previous version __________________________________ 20 Siemens AG SPR2-330.814.02.01.02 ARCADIS Avantic Page 3 of 20 Medical Solutions 02.05...
Required aids • Toolbox (including set of Allen keys) Configurations • The I.I. laser light localizer is available as an add-on for: - ARCADIS Avantic with 33 cm (13") I.I. ARCADIS Avantic SPR2-330.814.02.01.02 Siemens AG Page 4 of 20 02.05...
(e.g. cover of the basic unit or cover of the image intensifier) can be used in the protective conductor resistance report as the first measured value. Siemens AG SPR2-330.814.02.01.02 ARCADIS Avantic Page 5 of 20 Medical Solutions 02.05...
Page 6
The measuring procedure and the measuring device used (designation and serial num- ber) are also to be documented. Separate the page with the report from these instructions and file it in the "Reports" regis- ter in the system binder or log book. ARCADIS Avantic SPR2-330.814.02.01.02 Siemens AG Page 6 of 20 02.05...
Page 7
The protective conductor resistance must not exceed 0.2 ohms. The values determined in the repeat measurement, including the measuring points, must be recorded and assessed in the protective conductor resistance report. Siemens AG SPR2-330.814.02.01.02 ARCADIS Avantic Page 7 of 20 Medical Solutions 02.05 CS SD 24...
0.2 ohms is not exceeded. Laser light localizer option (Comment: Information applies to line-type lasers of laser class 1M, e.g., laser light localizers, I.I. / ARCADIS Avantic) CAUTION Laser emissions! This product contains lasers of the class 1M. (USA: Laser class Non-compliance can result in injury.
(1/Fig. 4 / p. 10)). • Pull out wires with plug that are wedged in at the side of the I.I. housing (see (1/Fig. 5 / p. 10)). Siemens AG SPR2-330.814.02.01.02 ARCADIS Avantic Page 9 of 20 Medical Solutions 02.05 CS SD 24...
Page 10
Installation and Settings Fig. 4: I.I._ring Fig. 5: Plug connector ARCADIS Avantic SPR2-330.814.02.01.02 Siemens AG Page 10 of 20 02.05 CS SD 24 Medical Solutions...
((1/Fig. 7 / p. 11) and wedge them with foam into the side of the I.I. housing ((1/Fig. 8 / p. 12). Fig. 6: I.I._1 Fig. 7: I.I._2 Siemens AG SPR2-330.814.02.01.02 ARCADIS Avantic Page 11 of 20 Medical Solutions 02.05 CS SD 24...
Page 12
For the mounting position of the laser light localizer, see ((1/Fig. 9 / p. 12)).For the position of the laser diodes see (2/Fig. 9 / p. 12) ARCADIS Avantic SPR2-330.814.02.01.02 Siemens AG Page 12 of 20 02.05 CS SD 24...
(AT THE SIDE OF THE LASER RING). Adjustment is done in the basic position (I.I. up, X-ray head down). Fig. 10: Adjusting the laser light localizer Siemens AG SPR2-330.814.02.01.02 ARCADIS Avantic Page 13 of 20 Medical Solutions 02.05 CS SD 24...
• Attach the included labels (1 und 2/Fig. 11 / p. 14) as shown (only attach for USA / Canada / UK). Fig. 11: I.I._Laser ARCADIS Avantic SPR2-330.814.02.01.02 Siemens AG Page 14 of 20 02.05 CS SD 24 Medical Solutions...
Check the function of the two keys to switch the laser light localizer on and off. • Switch the laser light localizer back on. The laser beams must go off automatically after approx. one minute. Siemens AG SPR2-330.814.02.01.02 ARCADIS Avantic Page 15 of 20 Medical Solutions 02.05 CS SD 24...
Installation and Settings Final Work Steps • File these instructions in Register 8, Service, in the system binder. ARCADIS Avantic SPR2-330.814.02.01.02 Siemens AG Page 16 of 20 02.05 CS SD 24 Medical Solutions...
Ensure that the I.I. laser light localizer is seated properly on the I.I. • Check the laser aperture for soiling and damage. • Check the alignment of the laser beams with the tube focus. Siemens AG SPR2-330.814.02.01.02 ARCADIS Avantic Page 17 of 20 Medical Solutions 02.05 CS SD 24...
Check the function of the two keys to switch the laser light localizer on and off. • Switch the laser light localizer back on. The laser beams must go off automatically af- ter approx. one minute. ARCADIS Avantic SPR2-330.814.02.01.02 Siemens AG Page 18 of 20 02.05...
Switch the laser light localizer back on. The laser beams must go off automatically af- ter approx. one minute. • Check the labels (see (Fig. 11 / p. 14)). Siemens AG SPR2-330.814.02.01.02 ARCADIS Avantic Page 19 of 20 Medical Solutions 02.05 CS SD 24...
Page 20
Changes since previous version 5Changes since previous version N/A; first edition ARCADIS Avantic SPR2-330.814.02.01.02 Siemens AG Page 20 of 20 02.05 CS SD 24 Medical Solutions...