Rear
Warning
Using this unit for medical purposes
The connectors on this unit are not isolated.
Do not connect any device other than one which conforms
to IEC 60601-1 standards.
When an information technology device or AV device
that uses an alternating current is connected, current
leakage may result in an electric shock to the patient or
operator.
If use of such a device is unavoidable, isolate its power
supply by connecting an isolation transformer, or by
connecting an isolator between the connecting cables.
After implementing these measures, confirm that the
reduced risk now conforms to IEC 60601-1 standards.
Caution
Do not come into contact with the terminals of the rear
panel connectors and patients at the same time.
Doing so may result in a generation of voltage that can be
harmful to patients if the unit is malfunctioning.
Always disconnect the power cord before connecting and
disconnecting connectors.
a
SDI input connectors A to D (BNC type)
Inputs SDI signals (3G SDI).
When [Ch1/Ch2 Rec Mode] is set to [Ch1 2K Rec] in
the [User Settings] screen - [Recording] tab, SDI
signals (3G SDI) are only input from SDI input
connector C.
Tips
• If SDI input images are not displayed, turn on the unit
first and make sure that startup is complete before
turning on the video input device.
• When receiving 4K signals, the four SDI inputs must be
synchronized.
• Embedded audio is not supported.
b
SDI output connectors A to D (BNC type)
Outputs SDI signals (3G SDI).
When [Ch1/Ch2 Rec Mode] is set to [Ch1 2K Rec] in
the [User Settings] screen - [Recording] tab, SDI
signals (3G SDI) are only output from SDI output
connector C.
c
HDMI output connector (Type A)
Outputs HDMI signals.
Note
We recommend using a Sony HDMI cable (not
supplied).