10. Disassemble the spacer bar and the unmounted frame. Use the bar as a straight edge
11. Wrap the Mylar spacing guide around the pipe so that the left edge is against the sensor
12. Realign left edge of the guide with the sensor edge mark. Line up both vertical edges of
13. Remove Mylar spacing guide and lay it out on a flat surface. Either measure the exact
14. Reinstall the spacing guide; its left edge abutting the sensors edge mark on the pipe and
FST020 IP65 NEMA 4X
Operating Instructions, 02/2012, A5E03086487-05
and, with one edge against the mounted frames tapered roller center and the other
crossing the dot you drew, draw a line crossing the dot (see "B" above). Set the bar
aside.
①
Trim material from inner edge if necessary
②
Mylar Spacing Guide
③
8cm (3-inch) Overlapping Edge
Figure 5-17
Wrapping the Mylar Spacing Guide around the pipe (End View)
edge mark (see "C" above). Arrange so that one end overlaps the other by at least three
inches. Trim to fit if necessary, but be sure not to trim at the overlapping end in order to
keep it square.
the guide and ensuring that it is snug around the pipe, mark along the overlapping edge.
distance half-way between the overlap edge and the mark at the overlap, or fold the
guide from the overlap edge to overlap mark and draw a line at the fold or halfway point.
①
Overlap Edge
②
Mark (or fold) exactly at half-way point
Figure 5-18
Finding the Halfway Distance
the overlapping edge in line with the dot (now a line) on the pipe (see "C). Tape it in this
position on the pipe. Take the second frame and place it against the edge of the guide
with its tapered roller centered on the center mark on the guide.
5.4 Installing the Sensor
③
Spacing Guide
④
Circumference
Connecting
45