Actions beginning with serial number 1409

Covers

Proceed as follows after the SIREMOBIL Iso-C has been moved to the respective room
and the brake has been enabled.

Cover of the X-ray unit

Remove the cover and reatttach them after installing the 3D option.

Cover of the horizontal carriage

Remove the cover and connector M3.X1.
Reattach the cover after installing the 3D option.

Cover of the cable module

Remove the cover and dispose of it properly, because this cover will no longer be
required.
Attach the new cover after installing the 3D option.

Centering disk, item no. 71 40 580

Fig. 1 Fully installed centering disk, beginning with serial number 1409
Attach the centering disk at the cable module (see Fig. 1).