Actions up to serial number 1408
Drilling and threading holes for installing the limit switches at the
cable module
Fig. 4 Drilling template limit switch
•
Use the drilling template for the limit switch (Fig. 4).
•
As seen from the operating console, two limit switches are installed to the left.
•
Mark the location of the holes (
•
Punch-mark the markings and keep the shavings from dropping into the cable module.
•
Continue to drill down to 15 mm using a 3.2 mm bit.
•
Counterbore the holes.
•
Cut the threads using an M4 thread tap (if required, use a drill with a left and right spiral
(clockwise and counterclockwise) or a tap wrench).
Siemens AG
Medical Solutions
Fig. 5 Thread for limit switch
/Fig. 4 and
/Fig. 5).
SPR2-230.814.01
Rev. 05
08.04
Page 3 of 14
SIREMOBIL Iso-C 3D
CS PS 24
2 - 3
2