Troubleshooting
10.4 Site Setup Data
10.4
Site Setup Data
This menu provides data pertaining to sensor characteristics and operation. Some menu
items are for technical support interpretation only.
Table 10- 4
fx Drive
N (burst length)
Ltn (mm/in)
Vf max
Vs max M/S
Vs min M/S
Empty
Samples/Cycle
Max Damping
Min Damping
HF
162
Site Setup Menu Items
Current Transmit drive code selected during
Initial Makeup. The drive code controls the sonic
transmit signal.
Transmit burst duration selected during Initial
Makeup. To change N count press <Right Ar-
row>. At equal sign enter numeric value (1 to 9
only).
Spacing distance between the sensors. It will be
in inches or millimeters, depending on default
units.
The flow velocity (in selected units) correspond-
ing to one whole cycle offset between upstream
and downstream receive signals.
Maximum Vs for current sensor spacing.
Minimum Vs for current sensor spacing.
Value of Empty Alarm Setting. The meter will
declare an empty status if signal strength drops
below this value.
Digital sampling rate.
Maximum signal damping. Use to average digital
data when an unstable condition occurs.
Minimum signal damping. Use to average digital
data when an unstable condition occurs.
Flow registration correction parameter.
Operating Instructions, 12/2014, A5E03086491-AC
FUE1010 IP65 NEMA 4X