Technical specifications
10.2 General technical specifications
Measuring ranges
(combination with O7)
Table 10- 23 ULTRAMAT 7: Gas inlet conditions
Gas inlet conditions
Sample gas pres-
Normal pressure (atmospheric pressure
sure
compensation)
Pressure loss between sample gas inlet and sample gas outlet
Sample gas flow
ULTRAMAT 7 standard version
ULTRAMAT 7 high-temperature
Sample gas temperature and humidity
ULTRAMAT 7 standard version
ULTRAMAT 7 high-temperature
Nominal values in the section "Determining the technical specifications"
1)
Table 10- 24 ULTRAMAT 7: Temperature of sample chamber and gas path
Temperature of sample chamber
and gas path
Gas path
Sample chamber
The high-temperature version of the analyzer module is not intended for installation in the rack-mounted device.
1)
102
C
H
2
4
SO
2
NO
N
O
2
NH
3
CO/NO
CO/NO
CO 0 ... 100/NO 0 ... 300 ppm
CO
/CO
2
ULTRAMAT 7
5 K above ambient temperature
5 K above ambient temperature
0 ... 300 ppm
0 ... 50 ppm
0 ... 100 ppm
0 ... 50 ppm
0 ... 100 ppm
0 ... 100 ppm
0 ... 100 ppm
500 to 1 500 hPa (absolute)
< 10 hPa at 1.5 l/min
18 to 90 l/h (0.3 to 1.5 l/min)
36 ... 90 l/h (0.6 ... 1.5 l/min)
0 to 50 °C at relative humidity < 90%, in reference to the
respective ambient temperature; avoid condensation in
gas path (moisture content results in measurement er-
rors)
0 to 65 °C, avoid condensation in the gas path (moisture
content results in measurement errors)
ULTRAMAT 7 high-temperature
≥ 60 °C with T
≥ 65 °C with T
65 °C
Compact Operating Instructions, 11/2019, A5E35134047-04
0 ... 100%
0 ... 100%
0 ... 3%
0 ... 100%
0 ... 100%
0 ... 10000 ppm
CO 0 ... 3000 ppm/NO 0 ... 3000 ppm
0 ... 100%
1)
1)
= 5 °C
a
= 20 °C
a
Devices in explosion-proof design
1)