2.7 Verify Calibration
For best results, verify the calibration daily.
Note: When not using the original transducer, calibrate the Model NK before use
(page 14).
1. Obtain the following values from the transducer calibration data sheet:
•
Shunt-cal output (or "shunt-cal factor") in mV/V
•
Calibration factor (or "full-scale output") in mV/V
•
Full-scale engineering units (or "capacity") in psig, lbs, etc.
2. Perform this calculation and record the value.
Note: Place a decimal point in the
the same place as set in the display.
3. With no load on the transducer, press and hold the
Compare the displayed value with the result of step 2. If the values are
not equal, adjust the
below.
Setup and Use
full-scale engineering units
potentiometer according to the procedure
fine span
value in the equation in
button.
shunt cal
11