- 3 1 1
Memory error
HP-IB message = "Power up state corrupt, set to
*RST state"
Local message = "Power
Description:
The signature check performed on the instruments
parameter store at power up, failed.
Possible reasons are:
parameter update
needs replacement
The instruments RAM is defective.
n
-311
Memory error
HP-IB message = "Pattern block corrupt, set to
Local message = "Pattern corrupt, set to 'RST"
Description:
The signature
pattern block at power up, failed.
Possible reasons are:
parameter update
-314
HP-IB message = "[rem/lot] Setting [no] corrupt,"
Local message = "[rem/lot] Setting [no] corrupt"
Description:
Save/Recall memory lost. Indicates that the
nonvolatile data saved
been lost.
HP 8114A Service Guide
state corrupl,, set to
check
performed on the instruments
the 'SAV command has
Troubleshooting 5-37