hit counter script

HP 8114A Service Manual page 179

Programmable pulse generator 100 v/2 a
Table of Contents

Advertisement

MODEL 8111A
Page
2-O. Delay
Add the following at the end of the "Variable Range" paragraph:
"No warning adjustment range: 0.00 ns to at least 85.9% of the period
Page 2-9. Double Pulse Delay
Heading "Minimum Period": delete 133.4 ns and insert:
"134 ns (approx 7.5 MHz; double pulses at 7.5 MHz simulates a 15 MHz signal)."
Heading "Range": substitute the following:
"Range: 20.0 ns to 999 ms (max value: period - width - 4ns).With Hi-Z source resistance, the
minimum range inceases to typically 30 ns, and the maximum decreases to typically
period - width - 10 ns.
Range minimum: 104% of Width
Range maximum: at least 85.9% of period -width (for periods >7.5 us the maximum is 95% of
Page 2- 11, Inhibit Input
Change threshold spec to "TTL compatible"
Section 3 Testing the HP 8114A
Pages 3-77, -20, -23, -27, -29, -30, -31, -32, -34, -37, -39, and -41:
Identify the connectors of the high power attenuator; the end connected to the 8
be marked "IN" and the other end "OUT"
Page 3-28:
Identify the connectors of the high power attenuator; the end connected to the DVM should be
marked "IN" and the other end "OUT"
Page 3-30:
The 50-ohm feedthrough should be at the DVM-end of the cable.
page l-4 and 2- 1 Environmental:
add errata:
Altitude up to 2000m
Installation Category II
Pollution Degree 2
Warning: To prevent electrical shock, disconnect the HP model 8114A from mains before cleaning.
Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not
attempt to clean internally.
2. September 1997
(95% can be achieved with periods >7.5 us)."
period -width)."
Page 3 of 6