hit counter script

Test 5.3: Delay Jitter; Test Specifications; Equipment Needed; Procedure - HP 8114A Service Manual

Programmable pulse generator 100 v/2 a
Table of Contents

Advertisement

Test 5.3: Delay Jitter

Test Specifications

Equipment Needed

Procedure

H P 81 l4A UNDER TEST
0.03% + 25 ps (0.05% + 25 ps for 50 ns < delay <
RMS-Jitter
100 ns)
Digitizing Oscilloscope with Accessories
All cables: 50 0, coaxial, BNC, 122cm (4ft)
High Power Attenuator
1. Connect HP 8114A to the Scope as shown:
Equipment Set-up for Delay Jitter Test
2. For calculating the RMS-jitter, the rise time of the reference signal
within a 1% amplitude window is required. If this value is not,
already measured in the Period Jitter test, then perform the first, 6
steps of the Period Jitter test.
HP 54 12
2
3
RIC
\
ADAPTOR ,
A D A P T O R
Testing the HP g114A

Hide quick links:

Advertisement

Table of Contents

Troubleshooting

loading

Table of Contents