Changing the device language (optional)
Up to two languages are possible. English is always the second language. The devices will always
start in the first language. Please notify Honeywell if you wish to change the first language installed
in the device. If English is selected as the first language, the device can only be operated in
English.
The design of displays and menus, for example, date and time format, depends on the current
parameter set and language. If the language is changed, all the data in the device will be renamed
or reformatted and displayed in the current language. Carry out the following steps in the "Home"
display to change the device language:
Step
Action
Select the "Globe". The alternative language is shown
1
below the symbol Confirm your selection.
The language has changed. The text beneath the
2
"Globe" shows the previous language as the new alternative.
Error display after start-up or for system errors
A system error may occur in very rare cases. If this happens, a prohibition sign will be displayed,
and the device will switch to emergency mode.
Examples of these errors include uploading an incorrect parameter set or internal connection errors.
This means that after or during the start-up process, for example, a one-way street sign will be
displayed, and only the basic functions for troubleshooting will be active; all other functions are
disabled.
Rectify it by transferring a correct parameter set to the device, for example, the
 Details
If the error cannot be rectified by the parameter set, please contact Honeywell.
Information for general use
Rev. M / 73023639
GASLAB Q2 DISPLAYS (OPERATE THE DEVICE)
Figure 7.15: Device not ready
Was the error caused by the parameter set?
default parameter set.
8.4.11
Restoring the parameter set to factory settings (importing data)