If the following methods do not produce the required success, contact Honeywell.
Rectification of the errors under points 1 – 4 (gas supply error)
The following action can be taken if an error occurs in the gas supply, for example abnormal loss of
calibration gas.
.
Step
Action: Gas supply error rectification
Complete a tightness test by checking the entire gas route (from the
1.
cylinder/pipeline to the device) using a gas detector.
Seal any leaks properly.
2.
Set the regulators and valves to the required values.
3.
Disconnect blocked lines from the switched-off device and flush the pipeline with
non-flammable gas to remove any blockages and deposits not exceeding the
4.
maximum pressure.
Rectification of the errors under point 5 (electrical supply and insulation errors)
The following action can be taken if an error occurs in the electrical power supply. The use of a
voltmeter is required to locate the error. If the device is not supplied with electricity or the supply
voltage/current is too low, the following effects may occur:
•
The interactive screen is dark, the Status LEDs are off.
•
The interactive screen is dark, the right Status LED flashes or is off.
Step
Action: Testing the electrical supply
Check all switchgear and fuses outside the device and replace any defective parts.
1.
Check the cable connection between the device and the electrical supply (no break in
the wiring). Ensure that a low-resistance electrical contact is established.
2.
Reconnect the cores if necessary.
Check that voltage is being supplied to the inlet point and that the voltage is not
impaired or reduced by overload, insulation errors or defective components. Remove
3.
or replace these components if necessary.
Information for general use
Rev. M / 73023639
POSSIBLE MALFUNCTIONS AND TROUBLESHOOTING
Contact
The
data
details are provided at the start of this document.
225
GasLab Q2