8.4.11. Restoring the parameter set to factory settings (importing data)

If the device no longer produces acceptable results due to several incorrect settings, it may be
necessary to restore an older parameter set or the factory settings.
This procedure deletes all current settings in the device and should only be used as a last
option in an emergency. If in doubt, contact Honeywell. Before doing so, save all data that
is still required on an independent system. A backup must also be available.
Cette procédure supprime tous les réglages actuels de l'appareil et ne doit être utilisée
qu'en dernier recours en cas d'urgence. En cas de doute, contactez Honeywell. Avant de le
faire, enregistrez toutes les données encore nécessaires sur un système indépendant. Une
sauvegarde doit également être disponible.
8.6 Using enSuite functions in device
Step
Work: Restore default setting in the device
Start the enSuite configuration and analysis software.
1.
Back up all the relevant device data to an external data medium or computer.
2.
Upload the backup parameter set created as described in section
exporting the parameter set
3.
(If the backup is still in your enSuite installation, this item can be omitted)
Connect to the device.
4.
Transfer the backed up (default) parameter set to the device.
5.
The device will restart and then continue to operate with the settings from the
6.
backup file.
Import data
You can import selected files (backups) using
the icon above the red arrow (see image on the
right) in the enSuite window. Click on the icon
and follow the dialog boxes.
Once the file has been imported, it will
appear in the navigation window under the
parameter tree of the device, as shown in the
example on the right.
Do not forget to adjust the calibration gas properties if you are no longer using the same
calibration gas cylinder as when the device was delivered.
N'oubliez pas d'ajuster les propriétés du gaz d'étalonnage si vous n'utilisez plus la même
bouteille de gaz d'étalonnage que lorsque l'appareil a été livré.
Information for general use
Rev. M / 73023639
CONFIGURATION AND ANALYSIS SOFTWARE ENSUITE
operation.
to your PC.