Identifier
Warning:
Battery low
CPU load high
Power supply off
Power supply on
Restart
...by user
...after run time error
System watch dog
Warning signal:
Internal warning
Alarm signal: Internal alert
Identifier
Security switch open
User management:
User profile ... login
Identifier
Notes on actions or
situations
Warning
Standard A/B failure
Information for general use
Rev. M / 73023639
POSSIBLE MALFUNCTIONS AND TROUBLESHOOTING
Cause of System SFB messages
Low battery
Notify Honeywell
The CPU capacity is fully used. Do not start any additional
functions such as reading archives.
The power supply was switched off.
The power supply was switched on.
The reason specified caused a restart.
The power down was not error-free, for example not all files
were closed. This signal remains until it is accepted.
The basic system has suffered an error. The device is no
longer functioning perfectly.
Rectification: Please contact Honeywell.
Cause of User SFB messages
The system reports that the security switch has been opened.
The named user profile has logged in
Cause of Q2 AFB messages
Notes on actions or situations that are currently being
performed or are pending, e.g. operational calibration. The
analysis mode was left. The message disappears automatically
as soon as the task is completed, and the instrument is back in
analysis mode.
Results of actions: (Notes and warnings)
Rectification: The note will disappear automatically. The
message will be retained until the next successful operational
calibration.
It may not be possible to calculate the end value.
Rectification: Please contact Honeywell.
229
GasLab Q2