6.3.
Decommissioning / Dismantling / Disposal
Please proceed in the following order if you take the measuring instrument out of operation and may
want to reuse it later at another location or dispose of it.
Step
Action: Decommissioning
Switch off the voltages (supply and communication) and secure them against being
1.
switched on again. Any further disassembly is prohibited without this prerequisite!
Shut down all gas streams and depressurize all gas inlets. Prevent the unintentional
2.
opening of the shut-off valves by suitable measures so that no danger can arise from
further disassembly.
The battery buffers data for approx. one year and may need to be replaced afterwards.
Data for example, archive data and time settings, will be lost if the battery is empty. The
device will no longer start correctly. You will then require a Honeywell service visit.
La batterie tamponne les données pendant environ un an et peut avoir besoin d'être remplacée par
la suite. Les données, par exemple les données d'archives et les réglages de l'heure, seront perdues
si la batterie est vide. L'appareil ne démarrera plus correctement. Vous devrez alors faire appel à un
service d'entretien Honeywell.
To complete dismantling or disassembly, continue the above steps with the following, note before:
If the device contains hazardous media e, g. toxic gases, they must be removed before its
dismantling. Prevent the release of potentially pollutant or dangerous substances.
Si l'appareil contient des fluides dangereux, par exemple des gaz toxiques, ceux-ci doivent être retirés
avant son démantèlement. Empêcher le rejet de substances potentiellement polluantes ou
dangereuses.
Step
Action: Continue decommissioning with dismantling
Allow the device to cool down to ambient temperature and disconnect all the gas
3.
tubing at the couplings. Immediately seal all outlets and inlets.
Open the connection box in a non-potentially explosive atmosphere and remove all the
cables by releasing them in the cable gland or cutting them if they are sealed. Dismantle
4.
the device mechanically.
If you want to use the device later in another location, please bring it into the condition
5a.
in which it was transported. Observe the instructions in section
Observe the following information on disposal if you no longer wish to use the device.
5b.
Information for general use
Rev. M / 73023639
COMMISSIONING AND DECOMMISSIONING