Only if all the above requirements are satisfied should you continue with the other points in this
section. Proceed as described below.
Download the firmware you require from
instromet.com/en/gas-analyzers
the files on your parameterization device. Caution when saving the file. It is
not possible to use files if there are special characters in the file path.
Check whether your enSuite version is still up to date and download the latest
version if necessary from
in the Software Downloads section and install it as described in section
8.2.1
Open the security switch in the device (if nessesary)
Establish the data connection to the device
(in this case, M0004 is connected).
Highlight the relevant device in the navigation window either under the
Devices branch or under the Connections branch.
Read out all archive data and save this data (to avoid data gaps in the
archives as far as possible) In the lower area of the navigation window, select
the action Readout archives and save the archives under the desired name.
Read out the current parameter set, save these data, as original
parameterization. See sections
Saving, and exporting the parameter set.
Read the parameter set again for conversion after the update if the current
parameterization is no longer compatible.
Check carefully using the read parameter set whether the update can be used
(see requirements).
Select action Software configuration in the bottom section of the navigation
window (see next figure).
Information for general use
Rev. M / 73023639
CONFIGURATION AND ANALYSIS SOFTWARE ENSUITE
Installing
enSuite. .
https://www.elster-
in the Software Downloads section and save
www.elster-instromet.com/en/software-downloads
8.2.6 Read-out parameterization
199
and
8.2.7
GasLab Q2