Digital outputs
There are four digital outputs (electrically isolated passive output circuits) in the connection box
(TB6-terminals, details see figure 5.6).
The output at TB6-5 is a breaker (NC, "normally closed"), the other outputs are makers (NO,
"normally open"). The maximum load per channel is 28.8 V DC/120 mA. The maximum pulse rate is
25 Hz.
Cable type refer to 
5.4.2 Cables (power supply / communication)
820ft. The figure also shows the appropriate terminals in the connection box. The wiring in the
drawing is only designed to illustrate the screening using an example.
In addition to the wiring, various enSuite settings are required before this interface
can be used. 
Information for general use
Rev. M / 73023639
GASLAB Q2 AFTER DELIVERY AND AT THE PLACE OF USE
Figure 5.10: Example connection of digital outputs
Changing the IP address basic network and I/O settings
8.4.5
the maximum length is 250m ≙