A small mirror will also help you to find the connection pins and check that it is

6.1.4. Additional conditions for fiscal metering

In the case of fiscal use, the conditions laid down in the relevant approval must be met. This is
ensured by a file in the device that adjusts the settings. For final commissioning, the calibration
switch or security switch (SSW) must be closed. You can attach a seal directly to the SSW (jumper)
to prevent or detect unauthorized
For details on the SSW, see Section
(SSW)
for the approval file see
approval).
The applicable regulations on sealing the device must also be observed during fiscal operation.
Make sure that you comply with these regulations.
Information for general use
Rev. M / 73023639
COMMISSIONING AND DECOMMISSIONING
Use angled pincers or pliers to connect the jumper.
You can view the setting by using devices displays and enSuite.
See chapter
7 GasLab Q2 Displays (Operate the Device)
and chapter
8 Configuration and analysis software enSuite
access.
Within the scope of fiscal metering, depending on the regulations of
your country, for the final commissioning the presence of an authorized
person (calibration officer), who carries out officially protected work on
gas-technical plants is required.
Opening and closing the hardware parameter guard
6.1.3
8.4.10 Fiscal parameters and optional using approval file (Select
correctly positioned.
82