5.1.

Storage

Store the GasLab Q2 in a clean, dry place. Prevent dirt getting inside the device. Prevent
material damage caused by the formation of condensation.
endroit propre et sec. Empêchez la saleté de pénétrer à l'intérieur de l'appareil. Évitez des
dommages matériels dûs à la formation de condensation.
The following applies to storage:
The ambient conditions must be maintained. Do not expose the Q2
temperatures below −25°C ≙ −13°F ≙ 248 K or above +55°C ≙ +131°F ≙ 328 K.
11.4 Notes on device characteristics and operating conditions
The packages may only be stored in enclosed rooms.
Avoid mechanical impacts during storage.
The storage time with a battery is up to one year,
2 years if the battery is then replaced.
If the device is transported in cold weather or extreme temperature fluctuations have occurred,
GasLab Q2 must be slowly returned to room temperature (temperature at the place of use) in order
to prevent damage caused by the formation of condensation.
Do not remove transport guards and other safety devices, for example, for gas
connections, before the device has been installed at its destination and if dirt can still
penetrate it.
protection, par exemple des raccords gaz, n'est pas autorisé avant l'installation finale de
l'appareil, et cela demeure vrai tant et aussi longtemps que la saleté peut pénétrer
l'appareil.
Storage is also crucial for devices that are to be reused after dismantling.
6.3Decommissioning / Dismantling / Disposal.
Removed devices should generally be stored in the same way as new devices.
We recommend replacing the battery with a new one (for safety reasons done by
Honeywell) if the device has not been supplied with power for more than one year.
After two years without a power supply, the battery charge will be less than 20% and
After three years, the battery will be completely discharged. Please contact
Information for general use
Rev. M / 73023639
GASLAB Q2 AFTER DELIVERY AND AT THE PLACE OF USE
Le retrait des dispositifs de protection de transport et des dispositif de
it must be replaced by Honeywell service personnel.
Honeywell if you wish to store the device for a longer period.
Gardez le GasLab Q2 dans un
56
GasLab Q2