These control and information lines are followed by the measurement and calculation values
(results).
The values displayed are only valid if the system is in analysis mode and has not
suffered an error. This means that no sensor alarm is active, and both "Operation"
If the value is monitored, this is also shown clearly by a display in black text.
The rule of thumb is: There is no alarm active if the status LED is permanently on.
The following is shown in the Q2 main display:
Displayed value
(abbreviation)
HsV (in large letters)
Rho (in large letters)
CO2 (in large letters)
Ws
MN
Calculated gas analysis with 10 components and the total of the C
Displayed value
(abbreviation)
CH
4
N
2
C
H
6
2
C
H
3
8
nC
H
4
10
iC
H
4
10
nC
H
5
12
iC
H
5
12
C
H
6
14
C
H
7
16
C
H
8
18
CH
C
+
2
C6+
Information for general use
Rev. M / 73023639
GASLAB Q2 DISPLAYS (OPERATE THE DEVICE)
For information about changing or extending this setting,
section
8.4 Changing existing device settings (Parameterization)
and "Step" read "Analysis".
Meaning (of the value abbreviation)/Physical
quantity
Superior calorific value; based on volume
Density at base conditions
Carbon dioxide
Superior Wobbe index
Methane number
Meaning (of the value abbreviation)/Physical
quantity
Methane
Nitrogen
Ethane
Propane
n-butane
Isobutane
n-pentane
Isopentane
Hexane
Total heptane
Total octane
Total of all hydrocarbons
∑ ethane + all higher hydrocarbons
Hexane Σ + all higher hydrocarbons
132
Displayed result...
... in the preset unit
...Dimensionless
+ components
2
Displayed
result...
... in mol%
GasLab Q2