(a) Setting range of NZ2GF2B-60TCTT4
The relation between the set value of CH Input range (address: 100H, 130H, 160H, 190H) and temperature
unit is as the following.
CH Input range (address: 100H, 130H, 160H,
190H)
1 to 99
100 to 199
200 to 299
Set CH Input range (address: 100H, 130H, 160H, 190H) according to the thermocouple type.
Temperature
Thermocouple type
measurement
range
0 to 1700
R
0 to 3000
-200.0 to 400.0
0.0 to 400.0
0 to 1300
0 to 500
0.0 to 500.0
K
0 to 800
0.0 to 800.0
-200.0 to 1300.0
0 to 1000
0.0 to 1000.0
0 to 2400
0.0 to 400.0
0 to 500
0.0 to 500.0
0 to 800
0.0 to 800.0
J
-200.0 to 1000.0
0 to 1200
0 to 1000
0.0 to 1000.0
0 to 1600
0 to 2100
-200 to 400
-200 to 200
-200.0 to 400.0
0 to 200
T
0 to 400
0.0 to 400.0
-300 to 400
0 to 700
0.0 to 700.0
Thermocouple is used. (No input from other analog modules
(such as an A/D converter module)) (1 to 199)
Other analog modules (such as an A/D converter module) are
used. (200 to 299)
Celsius ()
CH Input range
/Fahrenheit
Resolution
(address: 100H,
()/digit
130H, 160H, 190H)
1
1
1
105
0.1
38
0.1
36
1
2 (Default value)
1
11
0.1
40
1
12
0.1
41
0.1
49
1
100
0.1
130
1
101
0.1
37
1
13
0.1
42
1
14
0.1
43
0.1
50
1
3
1
102
0.1
131
1
103
1
104
1
4
1
21
0.1
39
1
19
1
20
0.1
45
1
110
1
109
0.1
132
Item
Output temperature unit is Celsius ().
Output temperature unit is Fahrenheit
().
Unit is digit.
Auto-setting at input range change
CH Upper limit setting
CH Lower limit setting
limiter, CH Process
limiter, CH Process
alarm upper lower limit
alarm lower lower limit
value, CH Process
value, CH Process
alarm upper upper limit
alarm lower upper limit
value
1700
0
3000
0
4000
-2000
4000
0
1300
0
500
0
5000
0
800
0
8000
0
13000
-2000
1000
0
10000
0
2400
0
4000
0
500
0
5000
0
800
0
8000
0
10000
-2000
1200
0
1000
0
10000
0
1600
0
2100
0
400
-200
200
-200
4000
-2000
200
0
400
0
4000
0
400
-300
700
0
7000
0
APPENDICES
A
*1
value
313